English
Language : 

S9295 Datasheet, PDF (4/4 Pages) Hamamatsu Corporation – Large area photodiode integrated with op amp and TE-cooler
Si photodiode with preamp S9295 series
Precautions for use
s ESD
S9295 series may be damaged or their performance may deteriorate by such factors as electro static discharge from the
human body, surge voltage from measurement equipment, leakage voltages from soldering irons and packing materials. As a
countermeasure against electro static discharge, the device, operator, work place and measuring jigs must all be set at the
same potential. The following precautions must be observed during use:
•To protect the device
strap or similar tools
tforogmroeulnedctrthoestoaptiecradtioscr,hsabrgoedywvhiaicha
accumulate on the operator or the
high impedance resistor (1 MΩ).
operator,s
clothes,
use
a
wrist
•A semiconductive sheet (1 MΩ to 10 MΩ) should be laid on both the work table and the floor in the work area.
•When soldering, use an electrically grounded soldering iron with an isolation resistance of more than 10 MΩ.
•For containers and packing, use of a conductive material or aluminum foil is effective. When using an antistatic material, use
one with a resistance of 0.1 MΩ/cm2 to 1 GΩ/cm2.
s Strength
Thermoelectrically-cooler devices may be damaged if subjected to shock, for example drop impact. Take sufficient care
when handling these devices.
s Lead forming
When forming the leads, take care not to apply excessive force to the lead sealing glass. Excessive force may impair the
hermetic sealing, possibly degrading the cooling capacity.
To form the leads, hold the roots of the leads securely with a pair of pliers and bend them.
s Heatsink
Use a heatsink with thermal resistance less than 1.3 °C/W. Apply thermal grease between the heatsink and detector
package, and then fasten them with the screws. Be careful not to give any excessive force or mechanical stress to the
detector package at this point.
s Wiring
• Be careful not to misconnect the plus and minus leads of the thermoelectric cooler or preamplifier. Supplying a voltage or
current while these connections are reversed may damage the device.
• The feedback resistor integrated into S9295 series is high so it is susceptible to external noise. Always ground the case
terminal when using S9295.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2004 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K., Solid State Division
1126-1 Ichino-cho, Higashi-ku, Hamamatsu City, 435-8558 Japan, Telephone: (81) 53-434-3311, Fax: (81) 53-434-5184, http://www.hamamatsu.com
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P.O.Box 6910, Bridgewater, N.J. 08807-0910, U.S.A., Telephone: (1) 908-231-0960, Fax: (1) 908-231-1218
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49) 08152-3750, Fax: (49) 08152-2658
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: 33-(1) 69 53 71 00, Fax: 33-(1) 69 53 71 10
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44) 1707-294888, Fax: (44) 1707-325777
North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171 41 Solna, Sweden, Telephone: (46) 8-509-031-00, Fax: (46) 8-509-031-01
Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39) 02-935-81-733, Fax: (39) 02-935-81-741
Cat. No. KSPD1064E01
4
Dec. 2004 DN