English
Language : 

L12161-07_15 Datasheet, PDF (3/4 Pages) Hamamatsu Corporation – 150 kV MICROFOCUS X-RAY SOURCE
X-RAY TUBE VOLTAGE / CURRENT OPERATION RANGE
TXPRB0021EB
600
500
LARGE FOCUS MODE
400
MIDDLE FOCUS MODE
300
200
SMALL FOCUS MODE
100
10
40 50 60 70 80 90 100 110 120 130 140 150
TUBE VOLTAGE (kV)
* The X-ray tube voltage guaranteed range is 40 kV to 150 kV.
* Operation is not guaranteed when the tube current is below 10 µA.
DIMENSIONAL OUTLINES (Unit: mm)
X-RAY TUBE UNIT L8122-01
3 × M3 DEPTH 5
43°
228 ± 2
X-RAY
BEAM
ANGLE
43°
X-RAY
FOCAL SPOT
4 × M6 DEPTH 8
(MOUNTING
SCREW HOLES)
COOLING
FAN
PROTECTIVE
EARTH
TERMINAL
COOLING FAN
CONTROL CONNECTOR
200 ± 0.5
11
100 ± 0.5
100 MIN.
140 ± 2
23
4 × M6 DEPTH 8
(MOUNTING
SCREW HOLES)
298 ± 2
X-RAY CONTROL UNIT C12163
POWER INTER LOCK X-RAY ON ERROR
X-RAY150 CONTROL UNIT
426 ± 2
482 ± 2
RS-232C
MONITOR
RS-232C
CONTROL
INTER LOCK 1
INTER LOCK 2
CONTROL
OUTPUT
COOLING FUN
POWER
AC INPUT
FUSL 14 .0/250V
40
AC POWER INPUT CONNECTOR
300 ± 2
TXPRA0014EB
100 MIN.
33 Max.
(CONNECTOR)
30
CONNECTION CABLE LENGTH: 5 m
TXPRA0015EA