English
Language : 

L12161-07_15 Datasheet, PDF (2/4 Pages) Hamamatsu Corporation – 150 kV MICROFOCUS X-RAY SOURCE
SPECIFICATIONS
GENERAL
Parameter
X-ray tube voltage setting range
X-ray tube current setting range
X-ray tube voltage operational range 1
X-ray tube current operational range 1
Small Focus Mode
Maximum output
Middle Focus Mode
Large Focus Mode
X-ray focal spot size
(Nominal value)
X-ray beam angle 2
Small Focus Mode
Middle Focus Mode
Large Focus Mode
Focus to object distance (FOD)
Rated output
Communication method
Description / Value
0 to 150
0 to 500
40 to 150
10 to 500
10
30
75
7 (5 µm at 4 W)
20
50
Approx. 43
Approx. 17
Continuous rating
Interface: RS-232C (9-pin D-sub connector)
X-RAY TUBE UNIT
Parameter
X-ray output window material / Thickness
Target naterial
Operating ambient temperature
Storage ambient temperature
Operating and storage humidity
Weight
Description / Value
Beryllium / 0.2
Tungsten
+10 to +40
0 to +50
20 to 85 (No condensation)
Approx. 13.5
X-RAY CONTROL UNIT
Parameter
Input voltage (AC)
Power consumption
Operating ambient temperature
Storage ambient temperature
Operating and storage humidity
Weight 3
Description / Value
Single phase 100 to 240 (50 Hz / 60 Hz)
Less than 220
+10 to +40
0 to +50
20 to 85 (No condensation)
Approx. 6
REGULATION AND STANDARDS
Parameter
RoHS directive
EMC
Safety
Description
EN 50581 Category 9
IEC/EN 61326-1
Emission limits: CISPR 11 Group 1 Class A
Immunity requirements: Table 2
IEC/EN 61010-1
CONTROL SOFTWARE 4
Parameter
Description
Applicable PC
PC / AT compatible
Applicable OS
Windows® XP, 7
Interface
RS-232C interface
NOTE: 1See the graph of the X-ray tube voltage / current operation range.
2Reference value: With 50 % of maximum X-ray emission.
3This weight includes the accessories of approx. 1.5 kg.
4The control software is provided as a sample software for the purpose of MFX operation.
The performance of the software is not guaranteed.
Unit
kV
µA
kV
µA
W
µm
degree
mm
—
—
Unit
µm
—
°C
°C
%
kg
Unit
V
W
°C
°C
%
kg
Unit
—
—
—
Unit
—
—
—
! PRECAUTIONS TO USE
· This microfocus X-ray source generates X-rays harmful to the human body. Use sufficient caution when handling the equipment to avoid
direct or inadvertent exposure to X-rays.
Install the X-ray source or the X-ray tube unit in an X-ray shielded cabinet or room equipped with safety interlock functions to prevent
accidental exposure to X-rays.
OPERATIONAL CAUTION
· This microfocus X-ray source generates X-rays and must therefore be used only under the supervision of qualified personnel.
· This microfocus X-ray source shall be used in compliance with health and safety regulations enforced in order to prevent health hazards
problems due to ionizing radiation.