English
Language : 

C7942CA-02 Datasheet, PDF (3/6 Pages) Hamamatsu Corporation – Acquire digital X-ray image in real time and 170 & 176 mm diagonal size
Flat panel sensor C7942CA-02, C7943CA-02
s Timing chart
To acquire images through an image grabber board, write parameters in the software program or parameter file by referring to
the following timing chart and description.
1 FRAME
Vsync+ (GRABBER)
Hsync+ (GRABBER)
1st ROW
2nd ROW
He
LAST ROW
DUMMY LINES
1st ROW
2nd ROW
All pulses are counted in the
“High” period of Hsync+
ENLARGED VIEW
Vsync+ (GRABBER)
Hsync+ (GRABBER)
Phe
Pclk+ (GRABBER)
1st ROW
2nd ROW
Pg Phe Phd Phe Phd
All pulses are counted at the riging edge of Pclk+.
The effective video output is only included in the “Phe” period.
KACCC0362EA
Parameter
C7942CA-02
1×1
2×2
He
Effective line
Dummy line
2344
56
1172
28
Phe
Effective pixel
Dummy pixel
2240
160
1120
80
Phd
526
1726
Pg
21
21
Note: “He” is the Hsync count. Phe, Phd and Pg are the Pclk count.
Pulse
4×4
586
14
560
40
2326
21
1×1
1220
28
1216
32
334
27
C7943CA-02
2×2
610
14
608
16
958
27
4×4
305
7
304
8
1270
27
s External trigger mode
To acquire images in external trigger mode, input an external trigger pulse as shown below. When the time Tvd has passed after
the rising edge of the external trigger pulse, synchronous signals and video signals are output.
From Tmin to Tmax
RECOMMENDATION: 50 % OF FRAME TIME
ExtTrgLemo
ExtTrgGrb
(TTL)
Tvd
Vsync+
(RS-422)
Hsync+, Pclk+ and Data 1-12 are the same as internal trigger mode.
· Tmin is defined as 1/Sf (int).
· Tmax is defined as the reciprocal of the minimum value of Sf (ext).
· Tvd=390 µs (C7942CA-02)
210 µs (C7943CA-02)
KACCC0369EA
3