English
Language : 

C9312SK-06 Datasheet, PDF (2/6 Pages) Hamamatsu Corporation – Flat panel sensor High-speed, High reliability, 110 kV (X-ray tube voltage) proof
Flat panel sensor C9312SK-06
s General ratings
Parameter
Pixel size
Photodiode area
Number of pixels
Number of active pixels
Readout
Video output
Output data rate
Synchronous signal (Vsync, Hsync, Pclk)
ExtTrgGrb, ExtTrgLemo, IntExt
Scintillator
Specification
50 × 50
124.8 × 115.2
2496 × 2304
2472 × 2184
Charge amplifier array
LVDS (differential) 12-bit dual port
33.333
LVDS (differential)
TTL
GOS
Unit
µm
mm
pixels
pixels
-
-
MHz
-
-
-
s Absolute maximum ratings (Ta=25 °C)
Parameter
Symbol
Value
Unit
Supply voltage for digital circuitry (+5 V)
D.vdd
+6.0
V
Supply voltage for analog circuitry (+5 V)
A.vdd
+6.0
V
Input voltage (ExtTrgGrb, ExtTrgLemo, IntExt)
A.vdd
+6.0
V
Operating temperature *1
Topr
0 to +35
°C
Storage temperature *1
Tstg
0 to +50
°C
*1: No condensation
s Specification (Ta=25 °C, A.vdd= 5.0 V, D.vdd= 5.0 V)
Parameter
Symbol
Min.
Typ.
Max.
Unit
Frame rate (single operation)
Sf (int)
7.6
8
-
frames/s
Frame rate external (single operation )
Noise (rms) *2
Sensitivity *3
Sf (ext)
-
Sf (int) to 1
-
N (rms)
-
700
-
S
16
20
-
frames/s
electrons
LSB/mR
Saturation charge
Csat
-
2.25
-
M electrons
Resolution *4
Reso
7.1
9
-
line pairs/mm
Dynamic range
Defect line *5
Output offset *6
-
-
3200
-
-
-
-
-
20
lines
-
-
65
200
LSB
*2: Internal trigger mode, single operation
*3: at 110 kVp without filter
*4: Spatial frequency at CTF=5 %
*5: A defective line is a vertical or horizontal line containing 4 or more consecutive pixels that produce less than 1/8 of the
average output from the surrounding pixels. In this case, one of the following conditions must be met.
Horizontal defective line: Defective pixels are continuously located from the opposite side of the shift register on the upper
side or lower side.
Vertical defective line : Defective pixels are continuously located, starting from Row 1091 toward Row 0 in the upper half
area.
Defective pixels are continuously located, starting from Row 1092 toward Row 2183 in the lower
half area.
Adjacent defective lines are not allowed in the vertical and horizontal directions.
*6: Average of all effective pixels in single operation at Sf (int)
Note: X-ray energy range is 20 k to 110 kVp.
s Resolution
(Ta=25 ˚C, A.vdd=5.0 V, D.vdd=5.0 V)
100
90
80
70
60
50
40
30
20
10
0
0 1 2 3 4 5 6 7 8 9 10
SPATIAL FREQUENCY (line pairs/mm)
KACCB0148EA