English
Language : 

C13053MA_15 Datasheet, PDF (1/9 Pages) Hamamatsu Corporation – Mini-spectrometer
Mini-spectrometer
FT series
C13053MA
Compact and thin, built-in high-sensitivity
CMOS image sensor
The mini-spectrometer FT (flat type) series is a polychromator provided in a compact, thin case that houses optical elements,
image sensor, and driver circuit. Spectrum data can be acquired by guiding measurement light into a mini-spectrometer
through an optical fiber and transferring the measured results to a PC via the USB connection. The incorporation of a high-
sensitivity CMOS image sensor maintains high sensitivity equivalent to that of a CCD and achieves low power consumption.
Moreover, the trigger function that can be also used for short-term integration enables spectroscopic measurement of pulse
emissions.
The product includes free evaluation software with functions for setting measurement conditions, acquiring and saving data,
drawing graphs, and so on. Furthermore, the DLL function specifications are disclosed, so users can create their original
measurement software programs.
Features
Applications
Compact, thin case
Sugar content and acidity detection of foods
High-sensitivity CMOS image sensor built in
(high sensitivity equivalent to that of a CCD)
With a trigger function
Plastic sorting
Thickness gauge
High throughput using quartz transmission grating
Highly accurate optical characteristics
External power supply not necessary (USB bus powered)
Installable in equipment
Stores wavelength conversion factor*1 in internal memory
*1: A conversion factor for converting the image sensor pixel number into a wavelength. A calculation factor for converting the A/D
converted count into the input light level is not provided.
Optical characteristics
Parameter
Specification
Unit
Spectral response range
500 to 1100
nm
Spectral resolution
Typ.
2.5
(FWHM)*2
Max.
3.5
nm
Wavelength reproducibility*3
-0.4 to +0.4
nm
Wavelength temperature dependence
Spectral stray light*2 *4
-0.04 to +0.04
-33 max.
nm/°C
dB
*2: When the slit in the table in “ Structure” is used. The spectral resolution depends on the slit.
*3: Measured under constant light input conditions
*4: The ratio of the count measured when an 800 nm light is input to the count measured when an 800 ± 40 nm light is input.
Electrical characteristics
Parameter
Specification
Unit
A/D conversion
16
bit
Integration time
11 to 100000
μs
Interface
USB 2.0
-
USB bus power current Typ.
220
consumption
Max.
250
mA
www.hamamatsu.com