SERIES
ULTRASONIC FLOWMETER <
DATA SHEET
F (Flexible Type)>
FLHâ¦3, FLWâ¦2, FLDâ¦1
This ï¬owmeter is a ï¬xed type ultrasonic ï¬owmeter based on
transit-time measuring method. Thanks to micro-processor
based electronics, the flowmeter can be easily configured
from the keyboard to speciï¬c applications, the ï¬owmeter is
suitable for liquid ï¬ow measurements for pipe size 50mm to
6000mm diameter.
The flowmeter is a compact and light-weight instrument
incorporating the latest electronics and high speed digital
signal processing technologies (32bit MPU), realizing high
performance and easy operation. This high performance
ï¬owmeter is capable of realizing a simultaneous 2-path or
2-pipe system.
FEATURES
1. Compact and light-weight
This ï¬owmeter is designed through use of the latest elec-
tronics and digital signal processing technologies.
2. Full variety of sensors
The ï¬owmeter can be used with various types of sensors
applicable for small to large pipe (ø50 to ø6000mm) and
low to high ï¬uid temperature (-40 to +200°C).
3. High accuracy
The ï¬owmeter is designed for high accuracy (better than
±1.0% of rate) by dynamic correction of fully-developed
ï¬ow proï¬le. Reynolds Number is calculated and a meter
factor (K) is automatically applied for best accuracy at all
ï¬ow velocities.
The adoption of new sound velocity measurement system
permits measurements of ï¬uids of unknown sound veloc-
ity with slightly affecting from ï¬uid temperature and pres-
sure.
4. Excellent resistance against aerated ï¬ow
Fuji' s unique ABM feature improves measurement reli-
ability for different ï¬ows like slurries, sludge, raw sewage
and bubble-contained ï¬ow (acceptable up to air bubbles
of 12% volume at 1m/s velocity).
5. Simultaneous 2-path or 2-pipe system
Average flow rate on 2 propagation paths can be mea-
sured simultaneously with less inï¬uence from changes in
time of ï¬ow proï¬le. And also ï¬ow measurement in two
separate pipes with one FLH3 can be realized for calculat-
ing functions of ï¬ow.
6. Excellent performance and easy operation
Large LCD and function keys allow easy conï¬guration and
trouble shooting.
â LCD with back light
â Easy mounting of sensor
â Trouble shooting
Flow transmitter (FLH)
Small sensor (FLW12)
SPECIFICATIONS
Fluid conditions
Measured ï¬ow : Liquid ï¬ow through which ultrasonic signal
can be transmitted (water, sea water, oil
and ï¬uid having unknown sound velocity)
Turbidity :
10000deg (mg /L) or less
State of ï¬ow : Turbulent or Imainor uniform ï¬ow well
grown up
Fluid temperature :
Small sensor
Middle sensor
-40 to +80°C
Large sensor
High-temperature sensor -40 to +200°C
Velocity range : -32 to 0 to +32m/s
Piping conditions
Pipe material : Carbon steel, ss, cast iron, polyvinyl chlo-
ride, FRP, asbestos, copper, aluminum,
acryl, etc.
Pipe size :
Small sensor
ø50 to ø400
Middle sensor
ø200 to ø1200
Large sensor
ø200 to ø6000
High-temperature sensor ø50 to ø400
Lining material : None, tar epoxy, mortar, rubber, or other
material with known sound velocity
Straight pipe length :
Upstream, 10D or more
Downstream, 5D or more (D= Inside diam-
eter of pipe)
Refer to JEMIS-032 for details.
JEMIS :
Japan Electric Measuring Instruments
Manufacturersâ Associationâs standard.
EDSX6-111e