EDS3-136 Datasheet, PDF (1/16 Pages) Fuji Electric – INFRARED GAS ANALYZER FOR STIC GAS | |||
|
INFRARED GAS ANALYZER FOR STIC GAS
DATA SHEET
ZSK
This infrared gas analyzer (ZSK) is composed of gas ana-
lyzer units, oxygen sensor and gas sampling equipment.
It is capable of continuous measuring concentration of
gases such as NOx, SO2, CO, CO2 and O2 contained in
the ï¬ue gas generated from various boilers or garbage
incinerators.
FEATURES
1. Gas concentrations of 5 components measurable
simultaneously and continuously
The infrared ray method has integrated the measurements
of NOx, SO2, CO and CO2 and the gas concentrations
of 5 components can be measured simultaneously and
continuously with the zirconia oxygen analyzer.
2. A high-sensitivity mass ï¬ow sensor is used in the
detector unit of infrared method.
Due to use of single beam system for measurement,
maintenance is easy and an excellent stability is ensured
for a long period of time.
3. Space-saving conï¬guration is available with the
maintenance from the front structure
The main unit is downsized by half of that of the conven-
tional product. The complete maintenance from the front
makes it easy to use.
4. The analyzer has a host of capabilities
Provided with a variety of functions such as O2 correction
output, average value output, auto calibration, auto range
selection, alarm, etc.
SPECIFICATIONS
Standard Speciï¬cations
⢠Measuring principle:
NOx, SO2, CO, CO2: Non-dispersive
infrared absorption (NDIR) method
O2; Zirconia system
⢠Measurable component and min./max. measuring range:
NOx : 0 to 200ppm/0 to 5000ppm
SO2 : 0 to 200ppm/0 to 5000ppm
CO : 0 to 200ppm/0 to 5000ppm
CO2 : 0 to 10%/0 to 20%
O2 : 0 to 10%/0 to 25%
⢠Measuring range:
Max. range ratio 1:10 (See code sym-
bols.)
⢠Warm-up time: 4 hours or less after power ON
⢠Analog output signal:
4 to 20mA DC simultaneous output for
each (Non-isolated or isolated: as speci-
ï¬ed by code symbols)
⢠Instantaneous value output
(NOx, SO2, CO, CO2, O2) 5 points
⢠Instantaneous O2 correction value
(NOx, SO2, CO) 3 points
* Provided with O2
⢠O2 correction average value
(NOx, SO2, CO) 3 points
* Provided with O2
⢠Allowable load resistance: 550Ω or
less (750Ω or less for insulation out-
put)
⢠Contact output:
(1) 1a contact for each (Contact capacity:
250V AC 2A, 30V DC 3A)
⢠Range identiï¬cation for each compo-
nent, analyzer unit error, calibration
error, during automatic calibration,
during maintenance
(2) 1c contact for each (Contact capacity:
250V AC 1A, 30V DC 1A)
⢠Instantaneous/concentration value
alarm for each component (H, L, HL
can be set).
EDS3-136
Date Nov. 14, 2007
|
▷ |