EDS3-136 Datasheet, PDF (1/16 Pages) Fuji Electric – INFRARED GAS ANALYZER FOR STIC GAS
INFRARED GAS ANALYZER FOR STIC GAS
DATA SHEET
ZSK
This infrared gas analyzer (ZSK) is composed of gas ana-
lyzer units, oxygen sensor and gas sampling equipment.
It is capable of continuous measuring concentration of
gases such as NOx, SO2, CO, CO2 and O2 contained in
the flue gas generated from various boilers or garbage
incinerators.
FEATURES
1. Gas concentrations of 5 components measurable
simultaneously and continuously
The infrared ray method has integrated the measurements
of NOx, SO2, CO and CO2 and the gas concentrations
of 5 components can be measured simultaneously and
continuously with the zirconia oxygen analyzer.
2. A high-sensitivity mass flow sensor is used in the
detector unit of infrared method.
Due to use of single beam system for measurement,
maintenance is easy and an excellent stability is ensured
for a long period of time.
3. Space-saving configuration is available with the
maintenance from the front structure
The main unit is downsized by half of that of the conven-
tional product. The complete maintenance from the front
makes it easy to use.
4. The analyzer has a host of capabilities
Provided with a variety of functions such as O2 correction
output, average value output, auto calibration, auto range
selection, alarm, etc.
SPECIFICATIONS
Standard Specifications
• Measuring principle:
NOx, SO2, CO, CO2: Non-dispersive
infrared absorption (NDIR) method
O2; Zirconia system
• Measurable component and min./max. measuring range:
NOx : 0 to 200ppm/0 to 5000ppm
SO2 : 0 to 200ppm/0 to 5000ppm
CO : 0 to 200ppm/0 to 5000ppm
CO2 : 0 to 10%/0 to 20%
O2 : 0 to 10%/0 to 25%
• Measuring range:
Max. range ratio 1:10 (See code sym-
bols.)
• Warm-up time: 4 hours or less after power ON
• Analog output signal:
4 to 20mA DC simultaneous output for
each (Non-isolated or isolated: as speci-
fied by code symbols)
• Instantaneous value output
(NOx, SO2, CO, CO2, O2) 5 points
• Instantaneous O2 correction value
(NOx, SO2, CO) 3 points
* Provided with O2
• O2 correction average value
(NOx, SO2, CO) 3 points
* Provided with O2
• Allowable load resistance: 550Ω or
less (750Ω or less for insulation out-
put)
• Contact output:
(1) 1a contact for each (Contact capacity:
250V AC 2A, 30V DC 3A)
• Range identification for each compo-
nent, analyzer unit error, calibration
error, during automatic calibration,
during maintenance
(2) 1c contact for each (Contact capacity:
250V AC 1A, 30V DC 1A)
• Instantaneous/concentration value
alarm for each component (H, L, HL
can be set).
EDS3-136
Date Nov. 14, 2007