3.6 Digital Communication via I²C-Bus
The SLS-1500 liquid flow meter has been designed
for use with the SCC1 interface cables. For special
applications, direct communication with the flow meter
via the standard I²C-interface is possible. The
physical interface consists of two bus lines, a data
line (SDA) and a clock line (SCL) which need to be
connected via pull-up resistors to the bus voltage of
the system.
These lines may be used on 3.3 V or 5.0 V levels with
a recommended clock frequency of 100 kHz. For the
detailed specifications of this I²C communication,
please refer to specific I²C Application Notes from
Sensirion.
4 Fluidic Connection
Table 3: Fluidic Specifications and Pressure Rating
Parameter
SLS-1500
Wetted materials:
 Internal sensor tube material
904L high-performance stainless steel
 Fitting material
PEEK
 Sealing material
Fluid connector ports (fittings)
PTFE
1/4-28 flat bottom
1/8” OD tubing (recommended: min 2mm ID)
Pressure drop (at 40 ml/min, H2O, 23°C)
2.15 mbar
Total internal volume
~70 l
For more information on the fluidic connection please find: “Application Note Sensor Ports and Tubing Connections” in the
Download Center on our homepage.
5 Mechanical Specifications
Table 4: Mechanical Specifications
Parameter
Largest dimensions
Total mass
Inner diameter sensor tube
Protection class
SLS-1500
59 x 31.5 x 18 mm
~40 g
1.5 mm
IP54
2 x M2.5 thread for mounting
904L high-performance
stainless steel
2 x 1/4-28 flat bottom
PEEK
All dimensions in mm
www.sensirion.com
Version 1 – February 2015
3/5