English
Language : 

SLS-1500 Datasheet, PDF (1/5 Pages) List of Unclassifed Manufacturers – Compact Flow Meter for Industrial Applications
SLS-1500 Liquid Flow Meter
Compact Flow Meter for Industrial Applications
 Liquid flow rates up to 40 ml/min
 20 ms response time
 High chemical compatibility
 Modular electrical interface
Product Summary
The SLS-1500 liquid flow meter enables precise, non-
invasive measurements of dynamic liquid flow rates up
to 40 ml/min bi-directionally. Excellent biocompatibility
is ensured by the exclusive use of high-performance
stainless steel, PTFE and PEEK for the wetted parts.
The flow path of the SLS-1500 liquid flow sensor is
formed by an especially thin-walled, straight tube which
assures excellent sensitivity.
Interface Options
Digital
- I2C-Bus
- RS485-Bus
- USB Cable
Analog
- Voltage Output (0-10 V)
- Additional operation
modes
For more information on communication, please refer
to page 2 of this document.
1 Sensing Performance (preliminary, subject to change)
Table 1: Performance of SLS-1500 (all data for medium H2O, 23°C, 1 barabs unless otherwise noted)
Parameter
SLS-1500
Unit
Full scale flow rate
40
ml/min
Sensor output limita
65
ml/min
Accuracy below full scale
(whichever error is larger)
10
% of measured value
0.25
% of full scale
Repeatability error from zero to full scale
(whichever error is larger)
1
0.025
% of measured value
% of full scale
Temperature coefficient
(additional error per °C; whichever is larger)
0.25
0.00625
% measured value / °C
% full scale / °C
Flow detection response time 63
20
ms
Response time on power-up
25
ms
Operating temperature
+5…+50
°C
Ambient storage temperatureb
-10…+60
°C
Operating pressure
12 (175)
bar (psi)
Proof pressure
25 (360)
bar (psi)
aFlow rate at which the sensor output saturates, see Section 2 for performance specification between full scale and saturation point
bNon-condensing, flow path empty
www.sensirion.com
Version 1 – February 2015
1/5