HD160C Technical Specifications (continued)
Data Sheet
Function
Range
Accuracy
DC Current
Ranges
100 µA, 1000 µA, 10mA, 100mA, 400mA, 2A (Auto/Manual ranging)
Resolution
0.01µA in 100 µA range
Accuracy
100 µA range
1000 µA to 400 mA ranges
2 A range
± (0.5% rdg + 10 dgts)
± (0.5% rdg + 5 dgts)
± (1.5% rdg + 10 dgts)
Input protection
2 A/1500 V fast blow ceramic fuse 8×65 mm on A input , FP700
Burden�v��o�l�t�a�g�e�
µA range of 1 mV/1 µA, mA range of 10 mV/1 mA, 2A range of 500 mV/1A, 500 mV max. (2 V max. on
1000 µA, 100 mA, 400 mA, 2 A ranges)
OL Protection
A Input (F 2 A/1500 V, size 8 x 65 mm IR fast blow ceramic)
AC�C��u�r�re��n�t�T�r�u�e��r�m��s
Ranges
100 µA, 1000 µA, 10 mA, 100 mA, 400 mA, 2 A
Resolution
0.01 µA in 100 µA range
Accuracy (45 Hz to 1kHz)
100 µA to 100 mA
400 mA
2A
± (1.5 % + 10 dgts)
± (2.0 % + 10 dgts)
± (2.5 % + 20 dgts)
Voltage�b��u�r�d�e�n��
see�D��C�C��u�r�re��n�t
Conversion type
True rms ac coupled 10 to 100% of range
Crest factor
≤3
OL protection
see DC Current
Resistance
Ranges
1000 Ω, 10 kΩ, 100 kΩ, 1000 kΩ, 10 MΩ, 40 MΩ
Resolution
0.1 Ω in 1000 Ω range
Accuracy
1000 Ω to 1000 kΩ ranges
10 MΩ range
40 MΩ range
± (0.5 % rdg + 8 dgts)
± (1.0 % rdg + 10 dgts)
± (2.0 % rdg + 10 dgts)
Overload protection, all ranges 1500 V dc or 1000 V ac rms
Continuity Test
Audible indication
Less than 40 Ω
Response time
100 ms
Overload protection
1500 V dc or 1000 V ac rms
Capacitance
Ranges
40 ηF, 400 ηF, 4 µF, 40 µF, 400 µF (3999 counts) (Auto/Manual ranging)
Resolution
0.01 ηF
Accuracy
40 ηF, 400 µF ranges
400 ηF to 40 µF ranges
±(3.0% rdg +10 dgts)
±(3.0% rdg +5 dgts)
Test voltage
<1V
Test�F�r�e�q�u��e�n�c�y��
1.3 Hz on 40 ηF to 40 µF ranges; 0.7 Hz on 400 µF range
Overload�p��r�o�t�e�c�t�i�o�n�
1500 V�d��c��o���r���1�0�0��0�V���a���c�r�m��s
Temperature
Ranges
-20 °C ~ 1300 °C (-4° F ~ 2372 °F)
3999 counts
Resolution
1 °C, 1 °F
Accuracy
-20 °C ~ 10 °C
10 °C ~ 200 °C
200 °C ~ 1300 °C
-4 °F ~ 50 °F
50° F ~ 400 °F
400 °F ~2372 °F
±(2.0% rdg + 4 °C)
±(1.0% rdg + 3 °C)
±(2.0% rdg + 2 °C)
±(2.0% rdg + 8 °F)
±(1.0% rdg + 6 °F)
±(2.0% rdg + 4 °F)
Overload protection
1500 V dc or 1000 V ac rms
(continued on next page)