English
Language : 

HD160C Datasheet, PDF (2/4 Pages) List of Unclassifed Manufacturers – HD160C Technical Specifications
HD160C Technical Specifications
Data Sheet
General Specifications
Display
Polarity Indication
Input overload indication
Low Battery Indication
Display Update Rate
Oper. Temp.
Storage Temp.
Altitude
Temperature coefficient
Fuse
Power
Auto Power-Down
Battery Life (typical)
Dimensions, without holster
(H x W x D)
Weight (incl. battery)
Accessories
Case material
Safety
EMC
4 digit LCD, 9999 counts, with annunciators, menu features and 41 segment bargraph.
Automatic
OL, -OL.
B; less than 50 hours battery life remain, accuracy is no longer guaranteed
2/sec, nominal; 20/sec for bargraph.
0 °C to +50 °C @ 0 to 75 % R.H.
-20 °C to 60 °C @ 0 to 80 % RH, battery removed
2000 meters - indoor use
0.1 x (spec. accuracy)/°C (0 °C to 18 °C and 28 °C to 50 °C)
F 2 A/1500 V fuse (8 mm x 65 mm), I.R. 30 kA – Amprobe p/n FP700
Standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22
Meter powers down after approximately 30 minutes of inactivity. Not in Min/Max function.
150 hours, alkaline. Backlight usage consumes extra power and will decrease battery life significantly.
Backlight auto-off after approx. 60 seconds.
200 x 102 x 59 mm (7.9 in x 4.0 in x 2.3 in)
642 g (22 oz)
Heavy Duty CAT IV 1000 V Test leads with threaded alligator clips (TL1500), battery (installed), hex
wrench (inside holster), Holster with Magne-Grip strap, Type K thermocouple probe (TP255), Tempera-
ture Adaptor (TA‑1A) and User Manual.
Reinforced, high-impact, fire retardant thermoplastic
Meets EN 61010-1 �C�A�T��I�V��1�0�0�0��V��(�1�5�0�0��V��D�C��m��a�x�)��C�l�a�s�s�I�I,�E��N��6�0�5�2��9�:I�P�6�7�
Meets EN 61326-1
Electrical Specifications (Accuracy at 23 °C ± 5 °C, <75 % RH, guaranteed for one year.)
Function
DC Volts
Ranges
Resolution
Accuracy
Input Impedance
CMRR
NMRR
OL Protection
Transient protection
AC Volts True rms
Ranges
Resolution
Accuracy
Input Impedance
Conversion type
Crest factor
OL Protection
Transient protection
Range
Accuracy
1000 mV, 10 V, 100 V, 1500 V
0.1 mV in 1000 mV range
±(0.1 % rdg + 5 dgt)
10 MΩ
>120 dB up to 1500 V dc
>60 dB at 50 or 60 Hz
1500 V dc or 1000 V ac rms.
12 kV impulse (1.2 µS/50 µS) based on EN 61010-1:2001 impulse requirement for at CAT IV 1000
V/1500V dc product. This product should not be used in installations where transients exceed 12 kV.
1000 mV, 10 V, 100 V, 1000 V
0.1 mV in 1000 mV range
1000 mV (45 Hz to 400 Hz)
10 V, 100 V (45 Hz to 500 Hz)
10 V, 100 V (500 Hz to 2 kHz)
1000 V (45 Hz to 1 kHz)
± (1.2 % rdg + 10 dgt)
± (1.2 % rdg + 10 dgt)
± (2.0 % rdg + 10 dgt)
± (2.0 % rdg + 10 dgt)
10 MΩ
True rms, ac coupled 5 % to 100 % of range
≤3
1500 V dc or 1000 V ac rms
12 kV impulse (1.2 µS/50 µS) based on EN 61010-1:2001 impulse requirement for at CAT IV 1000 V
product. This product should not be used in installations where transients exceed 12 kV
Amprobe® Test Tools