1.1 Calibration Field Information
The SLQ-QT500 Liquid Flow meter holds two
calibrations, one for isopropyl alcohol (IPA) and one
for water. Each calibration is stored on a separate
calibration field (CF):
 Calibration field 0: IPA (factory default)
 Calibration field 2: H2O
The default calibration field (i.e. the active calibration
field at power up) can be permanently changed via
I2C or RS485 commands. Alternatively, the default
calibration field can be changed using the USB-
RS485 Sensor Viewer which is part of the Flow Meter
Kit and also available in the download center on the
Sensirion liquid flow webpage.
www.sensirion.com/liquidflow-download
The standard calibrations cover the flow range
from -2500 µl/s to +2500 µl/s. Extended calibration
ranges for both IPA and water are available on
request.
1.2 Sensor Measurement Configuration
The default measurement configuration of the
SLQ-QT500 is optimized for the measurement of
short dispensing processes. For measuring
continuous flow rates, a different setting for the micro-
heater is preferred. The default setting can be
permanently changed via I2C or RS485 commands.
Alternatively, the default heater setting can be
changed using the USB-RS485 Sensor Viewer.
2 Specification Charts
Figure 1: Flow meter accuracy and repeatability across the flow range. Relative error in % of measured value (left column) and
absolute error in µl/s (right column) for IPA (top) and H2O (bottom)