NPC-100 Datasheet, PDF (2/4 Pages) List of Unclassifed Manufacturers – NovaSensor Disposable Medical Pressure Sensors
NPC-100 and NPC-120
Series Specifications
The NovaSensor NPC-100 and NPC-120 Series
pressure sensor are specifically designed for use in
disposable medical applications. The devices are
compensated and calibrated per the Association
for the Advancement of Medical Instrumentation
(AAMI) guidelines for industry acceptability. The
sensors integrate a high-performance, pressure
sensor die with temperature compensation
circuitry and gel protection in a small, low-cost
package.
The NPC-100 and NPC-120 Series are
manufactured in a class 1000 clean-room to
minimize possible sources of contamination. A
specially designed silicon micromachined sensing
element is used to meet or exceed all industry
requirements while minimizing assembly and
test cost for maximum customer value. Thick-film
laser-trimming is employed for final compensation
and calibration. Sensitivity is maintained to ±1%
and linearity is better than 1% in the physiological
operating pressure range.
The NPC-100 and NPC-120 Series are batch-
manufactured in ceramic plate form and shipped
as an intact array or in Tape and Reel for easy
customer automation. This assembly method
draws from well-established manufacturing
techniques used in the electronics industry in order
to produce a quality, high volume product.
0.0875 in to 0.0905 in
(2.22 mm to 2.30 mm)
0.049 in to 0.052 in
(1.24 mm to 1.32 mm)
0.189 in to 0.192 in
(4.80 mm to 4.88 mm)
0.236 in to 0.246 in
(5.99 mm to 6.25 mm)
Radius 0.005 in
(0.13 mm) maximum
0.142 in to 0.145 in
(3.61 mm to 3.68 mm)
Material is clear Polycarbonate.
0.285 in to 0.289 in
(7.24 mm to 7.34 mm)
NPC-100 Series pressure sensor port interface
0.0875 in to 0.0905 in
(2.22 mm to 2.30 mm)
0.049 in to 0.052 in
(1.24 mm to 1.32 mm)
0.236 in to 0.246 in
(5.99 mm to 6.25 mm)
0.124 in to 0.126 in
(3.14 mm to 3.20 mm)
Radius 0.005 in
(0.13 mm) maximum
0.142 in to 0.145 in
(3.61 mm to 3.68 mm)
Material is clear Polycarbonate.
0.285 in to 0.289 in
(7.24 mm to 7.34 mm)
NPC-120 Series pressure sensor port interface
Parameter
General
Pressure Ranges
Overpressure
Value
–30 to 300
125
Units
mmHg
psi
Electrical @ 72°F (22°C) Unless Otherwise Stated
Input Excitation
1 to 10
VDC
Dielectric Breakdown
10,000
VDC
Risk Current
2
µA
Input Impedance
1800 to 3300
Ω
Output Impedance
285 to 315
Ω
Environmental
Temperature
Compensated
15°C to 40°C
°C
Operating
15°C to 40°C
°C
Storage
–25°C to 70°C
°C
Humidity
10 to 90
%
Light Sensitivity
1
mmHg
Operating Product Life
168
hours
Shelf Life
3
years
Mechanical
Weight
<0.0044
lb
Media Interface
Medical grade, dielectric gel
Gel Tube Interface
Material
Polycarbonate
Notes
–0.58 psi to 5.8 psi
minimum
Calibrated for 6 VDC
5
Maximum (per AAMI), 5
(59° to 104°F)
(59° to 104°F)
(–13° to 158°F)
maximum (per AAMI BP22)
(<2 g)
Parameter
Units
Performance Parameters*
Offset
mmHg
Sensitivity
µV/V/mmHg
Calibration
mmHg
Symmetry
%
Linearity (–30 to 100 mmHg) mmHg
Linearity (100 to 200 mmHg) % output
Linearity (200 to 300 mmHg) % output
Thermal Coefficient Offset mmHg/°C
Thermal Coefficient Span %/°C
Frequency Response
Hz
Phase Shift
degrees
Offset Drift
mmHg/8 hrs
Minimum
–25
4.95
97.5
—
—
—
—
—
—
1200
—
—
Type
0
5
100
—
—
—
—
—
—
—
—
—
Maximum Notes
25
5.05
102.5
2
±5
1
6
1
6
1.5
6
±0.3
3
±0.1
3
—
5
5
5
1
4,5
* 1. All values measured at 6 VDC and 71.6°F (22°C) and after five second warm up unless
otherwise specified.
2. Output of sensor with no pressure applied and a 150 kW resistor shorted across
+IN to +OUT.
3. Over a temperature range of 59°F to 104°F (15°C to 40°C).
4. Normalized offset/bridge voltage—8 hours after 20 seconds. warm-up.
5. Previously qualified, not tested in production.
6. Deviation from straight line drawn through zero and 100 mmHg data points.
7. These sensors are compatible with ETO (ethylene oxide) sterilization. The user is
responsible for establishing and validating suitable process conditions to prevent damage to
the sensor.
0.160 in to 0.170 in
(4.06 mm to 4.32 mm)
–
Pressure
port
Dielectric gel
Vent hole
Polycarbonate
gel tube2
Contact pads
0.320 in +0.010/-0.002
(8.13 mm +0.25/-0.05) 1
0.109 in to 0.119 in
(2.77 mm to 3.02 mm)
0.017 in
1
(0.43 mm)
1. Tolerances reflect typical flaring during the singulation process.
NPC-100 Series package diagram
0.010 in
(0.25 mm)