English
Language : 

HP03S Datasheet, PDF (2/7 Pages) List of Unclassifed Manufacturers – Low voltage, low power consumption
Hope RF
2010-2-3
PIN Description
Pin Name
VSS
VDD
MCLK
XCLR
SDA
Pin Number
6
5
4
3
2
SCL
1
HP03 Series of calibrated sensor module
HP03S
Version: 1.3
Type
G
P
I
I
I/O
I
Function
power ground
power VCC
master clock(32k) input
ADC reset input (keep low when system is in idle state)
. I2C data input and output
I2C clock input
* XCLR is to reset the AD converter (active low). XCLR should be set to high only during AD
conversion phase(reading D1,D2), at all other states, such as reading calibration factors, this pin
should be kept low.
* The quality of the MCLK signal can significantly influence the current consumption of the
pressure module. To obtain minimum current, remember to supply good quality MCLK signal
Absolute Maximum Ratings
Parameter
Supply Voltage
Over pressure
Storage Temperature
Symbol
Min
VDD
-0.3
P
Tstg
-40
Max
Unit
4
V
15
Bar(abs)
125
°C
Recommended Operating Conditions
Parameter
Supply Voltage
Supply Current
Symbol Conditions Min
VDD
2.2
I
VDD=3V
during conversion
Stand by
Operating Pressure Range
P
Operating Temperature Range T
MCLK
T
Duty Cycle of MCLK
Serial Date Rate
SCL
300
-40
30
40%
Typ
3
500
1
25
32768
50%
Max
3.6
1100
85
35
60%
500
Unit
V
V
μΑ
μΑ
hpa (abs)
°C
KHz
%
KHz
Pressure and Temperature Output Characteristics
With the calibration data provided by the HP03 system, it should be able to reach
the following characteristics:
Parameter
Symbol Conditions Min
Typ
Max
Unit
Absolute Pressure
Accuracy
750-1100
HP03SA
@
-1.5
0~50
1.5
hpa
Absolute Pressure
Accuracy
750-1100
HP03SB
@
-3.0
0~50
3.0
hpa
Absolute Pressure
Accuracy
750-1100
HP03SA
@
-3.0
-20~60
3.0
hpa
Absolute Pressure
Accuracy
750-1100
HP03SB
@
-5.0
-20~60
5.0
hpa
Long Term Stability
12 month
2
hpa
Page 2 of 7