English
Language : 

205-TC Datasheet, PDF (17/22 Pages) List of Unclassifed Manufacturers – THERMOCOUPLE MINIATURE PANEL THERMOMETER
2. Connect the thermocouple calibrator output signal to
the thermocouple input connector P3 as shown in
Figure 4-2. Connect the negative (red) thermocouple
wire to connector P3, Pin 1. Connect the positive
thermocouple wire to connector P3, Pin 2. Be sure to
observe polarity and use the correct thermocouple
wire (Type J wire for Type J meter).
T/C CALIBRATOR
T/C WIRE
P3
0°C
P2
METER
P1
AC
Figure 4-2. Calibration a Thermocouple Calibrate
3. Connect the power to the meter. Allow 30 minute
warm up before calibration of 0.1° resolution meters.
Refer to the appropriate model number in Table 4-1
for the proper calibration points for your particular
meter.
20
Z
S
S1
MOUNTING
SCREW HOLES
CLAMP
CLAMP
SCREWS
CASE
2.677 +0.028/-.000
[68.00 +0.70/-0.00]
R 0.06 [1.5]
4 PLCS
0.874 +.012/-.000
[22.20 +0.30/-0.00]
PANEL THICKNESS
0.25 [6.4] MAX
0.03 [0.8] MIN
PANEL CUTOUT
PANEL CUTOUT
LENS
MOUNTING CLAMP
MOUNTING SCREW (2)
Figure 3-3. Installation Information
9