Section 3.3 Installation and Panel Mounting
Figure 3-3 shows the panel cutout dimensions, the dimensions
for the panel thickness, and the mounting clamp for mounting
the unit in a panel. Connections will be done after mounting
the unit.
1. Remove the lens by inserting a paper clip or a small flat
blade screwdriver into the rounded corner of the lens and lift
out.
2. Remove the two mounting screws from both sides of the
display. Pry both sides of the mounting clamp out and down
to remove it from the meter case.
3. Insert the meter into the panel cutout and reinstall the
mounting clamp to the bottom of the meter case from the
rear of the panel.
4. Re-install the mounting screws to secure the meter in the
panel. Do not over-tighten the screws.
5. Pop the lens back in, making sure the dull side of the lens
faces towards you.
4. Set the thermocouple calibrator to match the
thermocouple type of your meter. Adjust the
thermocouple calibrator to output the temperature
shown in the first line of column 2. Check to see that
the unit displays the temperature shown in column 3.
If not, adjust the zero potentiometer until it reads that
simulated value. For example, if the meter has a
"-JC2" in the model number, set the calibrator for
0.0°C and the unit display should show -0.2°C.
5. Adjust the thermocouple calibrator to output the
temperature shown in the second line of column 2.
Check to see that the unit displays the temperature
shown in column 2. If not, adjust the span pot until it
reads that simulated value. For example, if the meter
has a "-JC2" in the model number, set the calibrator for
170.0°C and the meter display should show 170.0°C.
6. After adjusting the pots, go back and verify both
temperature readings. Repeat steps 4 and 5 as
required.
8