SAFETY CONSIDERATIONS
This device is marked with the international Caution, risk of danger symbol. It is
important to read this manual before installing or commissioning this device as it
contains important information relating to Safety and EMC (Electromagnetic
Compatibility).
Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt
to operate the unit if damage is found.
This instrument is a DIN rail mount device. Installation of this instrument should
be done by Qualified personnel. In order to ensure safe operation, the following
instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be
included in the building installation as a disconnecting device. It shall be marked to indicate
this function, and it shall be in close proximity to the equipment within easy reach of the
operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth
wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International
Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains
supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
• The Protective Conductor must be connected for safety reasons. Check
that the power cable has the proper Earth wire, and it is properly connected. It is
not safe to operate this unit without the Protective Conductor Terminal
connected.
• Do not exceed voltage rating on the label located on the top of the instrument
housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
• Unit mounting should allow for adequate ventilation to ensure instrument does
not exceed operating temperature rating.
• Use electrical wires with adequate size to handle mechanical strain and power
requirements. Install without exposing bare wire outside the connector to
minimize electrical shock hazards.
EMC Considerations
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems
persist.
INSTALLATION
1. Unpack and verify that all equipment has been received. If there
are any questions about the shipment, contact the service
department nearest you.
2. Cut or punch a hole in the panel where you want the meter to go.
3. Remove the lens by inserting a paper clip or small flat screwdriver
into the rounded corner of the lens and lift out.
4. Remove the two mounting screws from both sides of display. Pry
both sides of the mounting clamp out and down to remove it from
the case.
5. Insert the meter into the panel cutout and reinstall the mounting
clamp from the rear.
6. Reinstall the mounting screws to secure the meter into the panel.
Do not overtighten the screws.
7. Pop the lens back in, making sure the dull side of the lens faces
towards you.
PANEL CUTOUT
CASE
LENS
MOUNTING CLAMP
MOUNTING SCREW (2)
PANEL CUTOUT
2.677 +0.028/-.000
[68.00 +0.70/-0.00]
R 0.06 [1.5]
4 PLCS
0.874 +.012/-.000
[22.20 +0.30/-0.00]
PANEL THICKNESS
0.25 [6.4] MAX
0.03 [0.8] MIN
DIMENSIONS
0.945 [24.00]
0.197
[5.00]
2.835
[72.00]
4.134
[105.00]
4.60
[116.8]
0.858 [21.80]
2.661
[67.60]
SIDE VIEW
TOP VIEW
All dimensions are in inches [millimeters].
REMOVING BOARD FROM CASE
PULL TAB HERE
PC BOARD
PUSH BOARD ASS'Y FROM HERE
1 23 4 5 6 7
P1
REAR OF CASE
ELECTRICAL CONNECTIONS
POWER AND SIGNAL INPUT
~ AC POWER
= DC POWER
P1 205-E
205-P & 205-A 205-P & 205-A
1 SIG HI (+S)
SIG HI (+S)
SIG HI (+S)
2 SIG LO (-S)
SIG LO (-S)
SIG LO (-S)
3 + Exc
Hold
Hold
4 - Exc
Digital Return
Digital Return
5 AC Earth GND (©) AC Earth GND ( ) -DC (-)
6 AC Neutral (N)
AC Neutral (N)
+ DC (+)
7 AC Line (L)
AC Line (L)
No Connection (N/C)
WIRING
AC POWER WIRING
Connect the ac main power connections as shown below.
EARTH
NEUTRAL
LINE
GREEN WIRE
N
L
SWITCH
FUSE
Check for proper Earth grounding in the
power distribution system (single phase).
5
6
7
Screws
underneath
P1
ORANGE
CONNECTOR
CAUTION: Do not connect ac power to your meter until you
have completed all input and output connections. Failure to
do so may result in injury! This device must only be installed
electrically by specially trained electrician with
corresponding qualifications. The main power input to the
unit as well as the AC input signal to be measured must
agree with the wiring instruction.
The meter is factory set to the power specified by the customer at the
time of ordering. The voltage is printed on the label under Input
power.
To connect power to the unit, attach wires, as shown from the ~AC
power cord to the orange 3-pin female connector that is supplied with
the meter. Plug the female connector into the orange P1 (male)
connector at the rear of the meter.
DC POWER WIRING
The meter can be
No Connection N/C
configured at the
factory to operate on DC
External
Power
+ DC
Supply – DC
non-isolated power. The
DC voltage option is
specified at the time of
ordering. An AC powered meter
cannot be converted to a DC
5
6
7
powered meter or vice versa.
Screws
underneath
P1
ORANGE
CONNECTOR
To connect power to the unit, attach wires, as shown, to the orange 3-
pin female connector that is supplied with the meter. Plug the female
connector into the orange P1 (male) connector at the rear of the
meter.
SIGNAL INPUT WIRING 205-E
Current
Transmitter
– + + EXC
Sig Hi
Sig Lo
– EXC
1
2
P1
3
4
Screws
underneath
SIGNAL INPUT WIRING 205-P & 205-A
Switch
Hold
Digital Return
–
Current
OR +
Voltage ~
–
+
DC Current
DC Voltage
1
2
P1
3
4
Screws
underneath
RANGE SELECTION
CURRENT / VOLTAGE INPUT 205-P & -E
4-20 mA
0-0.2 V dc
1-5 V dc
0-10 V dc
0-100 V dc
VOLTAGE INPUT RANGE 205-A
±199.9 mV dc
±1.999 V dc
±19.99 V dc
±100.0 V dc
EXCITATION VOLTAGE 205-E ONLY
10 V, (standard internal reference)
24 V, (standard internal reference)
10 V, (reference based on excitation voltage)
S3
D, E
E
A
B
C
S3
-
A
B
C
S2
S4
B, D
A
B, D
-
A, C
A
Voltage operation was configured per customer order. 115 and
230 Vac operation use the same transformer and may be reversed
with simple wiring installation.
CAUTION: Disconnect power before attempting to convert or
remove the unit.
You must remove the printed circuit board from the case in order to
change from 115 Vac to 230 Vac (and vice versa). These changes
must be performed by a qualified technician in order to avoid damage
TRANSFORMER # AC VOLTAGE
48131
115 Vac
48131
230 Vac
INSTALL
W1, W3
W2
REMOVE
W2
W1, W3
CALIBRATION / 205-P & 205-E
1. Apply the minimum current or voltage to the signal input.
Adjust the zero pot for the minimum display reading ±1 count.
2. Apply the maximum current or voltage to the signal input.
Adjust the span pot for the maximum display reading ±1 count.
3. Repeat steps 1 and 2 until ±1 count desired reading. Reinstall the
lens.
CALIBRATION / 205-A
1. Apply a voltage equivalent to 95% of the range selected to the
signal input.
2. Adjust the span pot for a reading of 1900 ±1 count.
3. Reinstall the lens.
ZERO
Decimal
Point
S1
SPAN
1.999
A
Z
S
19.99
B
A
B
S1 C
S1
199.9
1999
C
Store jumper
above A