MDS-065APS15-B Datasheet, PDF (8/10 Pages) Delta Electronics, Inc. – MDS Medical AC-DC Open Frame
T``E`CHNICAL DATASHEET
MDS Medical AC-DC Open Frame
15Volt, 63Watt / MDS-065APS15 B
Power De-rating
Output Power
100%
50%
0%
-10°C
50°C
Ambient
70°C Temperature
Fig. 3
Functions
Start-up Time
The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied.
Rise Time
The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value.
Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time
required for the output to reach 95% of its set value, after the input voltage is removed.
■ Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
8
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)