English
Language : 

MDS-200ADB12AA Datasheet, PDF (5/11 Pages) Delta Electronics, Inc. – Medical AC-DC Enclosed Power Supply
Medical AC-DC Enclosed Power Supply
12V 200W 2”x4” High Density / MDS-200ADB12 AA
Protections
Overvoltage (max)
Overload / Overcurrent (max)
Over Temperature
Short Circuit
125% of rated voltage, Latch Mode
Main output 160% of ratedcurrent
Hiccup Mode(Non-Latching, Auto-Recovery)
Latch Mode
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Reliability
MTBF(Minimum) at 115Vac, 110W, 35°C, Convection Air Flow
Operating life at 115Vac, 110W, ambient 25 °C, Convection Air Flow
800 kHrs based on Telecordia SR-332
26,280Hrs
Safety Standards / Directives
Medical Safety
IEC60601-12nd ,3rd and 3rd +A1 edition CB report
TUV EN60601-1:2006
UL60601-1+CAN/CSA 60601-1: (Ed.3.2005)
ITE Safety
CE
Environmental
IEC60950-1 CB report
TUV60950-1
UL60950-1+CAN/CSA60950-1
GB4943.1-2011, GB9254-2008, GB17625.1-2003
MDD Directive 93/42/EEC
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
Input to/Output (2XMOPP) 4000 Vac
Input to/Ground(1XMOPP) 1500Vac1)
Output to/Ground(1XMOPP) 1500Vac ( Type BF application rated)
1) PSU can support PoE applications with Primary to FG 2500Vac test.
5
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)