English
Language : 

MDS-200ADB12AA Datasheet, PDF (2/11 Pages) Delta Electronics, Inc. – Medical AC-DC Enclosed Power Supply
Medical AC-DC Enclosed Power Supply
12V 200W 2”x4” High Density / MDS-200ADB12 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max)
Input Surge Voltage (max)
Full load Efficiency (typ.)
Standby Power (max)
Inrush Current (max)
Input-PE(protective earth) leakage current(max)
Output-PE(protective earth) leakage current for Type BF application (max)
Power Factor (min)
1) NC: normal condition, SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
2.5A
300Vac for 100ms
91% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
0.5W @ 115Vac/60Hz, 230Vac/50Hz
60A@230Vac, cold start
0.1mA @ NC, 0.3mA @ SFC1)
0.1mA @ NC, 0.5mA @ SFC 1)
0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
Normal Condition 18.6uA
Single Fault Condition 45.6uA
Output-PE Leakage Current for Type BF application
Normal Condition 39.2uA
Single Fault Condition 44.1uA
264Vac/60Hz(Typ) Delta Limit
44.8uA
100uA max
151.9uA
300uA max
82.2uA
128.1uA
100uA max
500uA max
IEC60601-1 Limit
5000uA max
10000uA max
100uA max
500uA max
100.000
90.000
80.000
70.000
60.000
115Vac
230Vac
50.000
25%
50%
75%
Output Load (200W)
100%
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)