MDS-300APB48-AA Datasheet, PDF (3/12 Pages) Delta Electronics, Inc. – Medical AC-DC Open Frame Power Supply
Medical AC-DC Open Frame Power Supply
48V 300W with 5V/0.5A Standby / MDS-300APB48 AA
Output Ratings / Characteristics
Nominal Output Voltage (Vrated)
Total Regulation
Output Power
Line Regulation (max)
Load Regulation (max)
Ripple & Noise (typ.)
Start-up Time (max)
Hold-up Time (min)
Dynamic Response (Overshoot & Undershoot O/P Voltage)
Capacitive load (max)
Rise time (max)
Remote Sense
48V
± 3%
300W 10CFM air, 240W convection air
±0.5%
±1%
1% pk-pk Vrated@300W load (Refer to Fig. 2 & Fig. 3)
2000ms @ 115Vac
10ms @ 100Vac and 115Vac, with 300W load
20ms @ 100Vac and 115Vac, with 190W load
± 5% @ with 50-100% load change
1500uF
100ms
Up to 500mV compensation for voltage drop across
external wire connections to load.
Short and reverse connection protected.
(a) 115V (measured value=392mV)
(b) 230V(measured value=384mV)
Fig.2 Ripple & Noise example, 20MHz BW
Output Ratings / Characteristics—Standby Output
Nominal Output Voltage of standby output (Vrated)
Total Regulation of standby output
Ripple & Noise of standby output
5V
± 3%
100mV max (Refer to Fig. 3)
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 13)