MDS-300APB48-AA Datasheet, PDF (10/12 Pages) Delta Electronics, Inc. – Medical AC-DC Open Frame Power Supply
Medical AC-DC Open Frame Power Supply
48V 300W with 5V/0.5A Standby / MDS-300APB48 AA
Dynamic Response (48 Volt Output)
The power supply output voltage will remain within ±5% of its
steady state value, when subjected to a dynamic load 50 to 100%
of its rated current.
■ 50 to 100% Load
Overload & Over current Protections
The power supply’s Overload (OLP) and Over current (OCP)
Protections will be activated before the 48 volt outputs’ current
exceeds 160% of IO (Max load). Upon such an occurrence, VO will
start to drop. Once the power supply has reached its maximum
power limit, the protection will be activated and the power supply
will go into “Hiccup mode” (Auto-Recovery). The power supply will
recover once the fault condition causing the OLP and OCP is
removed and IO is back within the specified limit. For the standby
output, the performance is the same, with the exception that the
protection will be activated before the output current exceeds
1.06 amps.
Inrush Current
Inrush current is input current that occurs when the input voltage
is first applied. For AC input voltages, the maximum peak value of
inrush current will occur during the first half cycle of the applied
AC voltage. This peak value decreases exponentially during
subsequent cycles of AC voltage.
Overvoltage Protection
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 5 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
Additionally, if the 48 Volts’ Iout is >100%; but <160% for a
prolonged period of time (depending on the load), the Over
Temperature Protection (OTP) will be activated due to high
temperature on critical components. The power supply will then
go into latch mode.
Short Circuit Protection
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
Over Temperature Protection
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating temperature condition at 100%
load, the power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into latch mode until the input
voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature
.
10
All parameters are specified at 25°C ambient unless otherwise noted.