Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current(max)
Input Surge Voltage (max)
Full load Efficiency (typ.)
Standby Power (max)
Inrush Current (max)
Input-PE(protective earth) leakage current(max)
Output-PE(protective earth) leakage current for Type BF application (max)
Power Factor (min)
1) NC: normal condition, SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
2.5A
300Vac for 100ms
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
0.5W @ 115Vac/60Hz, 230Vac/50Hz
60A@230Vac, cold start
0.1mA @ NC, 0.3mA @ SFC1)
0.1mA @ NC, 0.5mA @ SFC 1)
0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
Normal Condition 19.5uA
Single Fault Condition 44.8uA
Output-PE Leakage Current for Type BF application
Normal Condition 38.3uA
Single Fault Condition 43.4uA
264Vac/60Hz(Typ) Delta Limit
45.0uA
100uA max
153.0uA
300uA max
81.5uA
129.4uA
100uA max
500uA max
IEC60601-1 Limit
5000uA max
10000uA max
100uA max
500uA max
100
90
115Vac
80
230Vac
70
60
50
25%
50%
75%
Output Load
Fig.1 Efficiency versus output load
100%
2
All parameters are specified at 25°C ambient unless otherwise noted.