8420-51 Datasheet, PDF (7/8 Pages) ASM GmbH – MEMORY HiLOGGER | |||
|
◁ |
7
â Product Specifications
Functional Specifications
Display
5.7-inch STN Color LCD (240 Ã 320 dots). Displays either waveforms
only, waveforms and numerical values simultaneously, numerical
values only or calculation results and waveforms simultaneously.
Japanese/English language selectable. The most recent 16 MB of
data (with one-channel recording) can be displayed by back scrolling.
Waveform compression
and magnification
Time-axis: 1/2/5/10/20/30 seconds, 1/2/5/10/20/30 minutes,
1/2/5/10/12 hours, 1 day/division
Voltage-axis zoom: Ã100, Ã50, Ã20, Ã10, Ã5, Ã2, Ã1, Ã1/2
Search function Event marks can be searched
Calculation
functions
Waveform parameter calculations: Up to four simultaneous calculations
are supported: average, peak, maximum, minimum and effective values,
timing of maximum and minimum values, and waveform processing
calculation: displays sums and differences between channel waveforms
Communication
functions
(Controlled by PC)
â Data can be collected in real time using the optional LOGGER
COMMUNICATOR 9334 (data collection software application).
â Data in internal memory*1 and in the PC Card can be
transferred by the FTP server function.
*1 Data in internal memory can be transferred only when not measuring.
â Remote control by communications commands (separate
programming required)
â Remote control by HTTP server function (with no special software
other than a common Web browser, HiLOGGER settings, data acquisition and
screen monitoring can be performed)
Communication
functions
(Sending to PC)
â Automatic data sending by FTP client function (periodically
during measurement or after measuring, a ï¬le automatically saved to PC Card is
automatically sent to an FTP server on the internal network or on a remote PC)
â Automatic e-mail sending (upon trigger start/stop, trigger warning,
power outage recovery, memory or card full status, an e-mail notiï¬cation is sent to
a PC on the local network or to a remote PC through an SMTP mail server)
Communication
functions (PPP)
To communicate with a modem-equipped PC through the public
telephone circuit, connect a standard modem to the RS-232C
terminal using the RS-232C CABLE 9721.
Miscellaneous
Waveform scroll, cursor measurement, scaling, automatic saving
(binary/text selectable), start condition retention, settings saving, comment
entry, event marking (for search), automatic setup, saves the most recent
16 MB of data (with one-channel recording) in internal memory
Trigger Functions
Trigger source Analog input: all channels, Pulse totalizer inputs: P1 to P4, Logic inputs
(conditions can be set for LI-1 to LI-16 (using the DIGITAL I/O UNIT 8993), External trigger, Time
each channel)
trigger, Logical Product of each trigger source (AND), Logical Sum (OR)
External trigger
Active low, valid pulse width H period 2.5 ms or more, L period
2.5 ms or more (with external trigger ï¬lter ON)
Trigger timing Start, Stop and Start/Stop (separate trigger conditions can be set to start and stop)
Trigger type
(analog, pulse)
Level: Triggers when rising or falling through preset level
Window: Triggers when entering or exiting range defined by
preset upper and lower limit values
Trigger level resolution 0.5% f.s. (f.s. = 10 divisions)
Trigger type (logic) Pattern trigger by 1, 0 and à (don't care) (with the DIGITAL I/O UNIT 8993)
Miscellaneous
Pre-trigger (records period before trigger, can be set for real-time saving),
Trigger output (open collector, 5V output, active low, at least 100 ms pulse
width), Trigger mode (single, continuous)
Thermocouple Inputs
(accuracy specified @23 ±5ËC, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Screw-type terminal block (recommended minimum wire diameter 0.32 mm*2),
Input Terminals/
removable, supplied terminal block cover (all channel terminals are isolated from
each other and chassis), Input impedance: 1MΩ (850kΩ when open-circuit polling is
Impedance
enabled)
*2 Recommended wire: 0.14 to 1.5 mm2 single strand, or 0.14 to 1.0 mm2 twisted multi strand.
To connect smaller thermocouple wire, use the TERMINAL UNIT 9329.
Setting range
*3 Upper and lower limits depend on the
100ËC f.s. : -100 to 100ËC*3 (0.01ËC resolution)
500ËC f.s. : -200 to 500ËC*3 (0.1ËC resolution)
measurement input range of each sensor 2000ËC f.s. : -200 to 2000ËC*3 (0.5ËC resolution)
Thermocouples
are JIS C 1602-1995 compliant
(except type W)
K: -200 to 1350ËC, E: -200 to 1000ËC, J: -200 to 1200ËC, T: -200 to 400ËC,
N: -200 to 1300ËC, R: 400 to 1700ËC, S: 400 to 1700ËC, B: 400 to 1800ËC,
W (WRe5-26): 400 to 2000ËC
Note: Types W, R and S can measure from 0ËC, but measurements in this range are not
guaranteed Standard. Reference contact: internally and externally switchable
Accuracy
Types K, E, J, T, N: ±0.05% f.s. ±1ËC
Types R, S, B, W: ±0.05% f.s. ±2ËC (applicable at 400ËC and above)
Standard reference contact accuracy: ±1ËC
(with internal compensation, add to measurement accuracy)
Max. allowable input
60 V DC (maximum voltage between input terminals that does not cause
damage)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
Platinum resistance temperature sensor inputs
(only in 8420-51/8421-51, accuracy specified @23 ±5ËC, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Input terminal, Common thermocouple application, detection current: 1 mA,
detection current Input Resistance: 2 MΩ
Setting range 100ËC f.s. : -100 to 100ËC*4 (0.01ËC resolution)
*4 Upper and lower limits depend
on the measurement input range
of each sensor
500ËC f.s. : -200 to 500ËC*4 (0.1ËC resolution)
2000ËC f.s. : -200 to 2000ËC*4 (0.5ËC resolution)
Resistance
Pt100: -200 to 800ËC, JPt100: -200 to 500ËC, Connection:
temperature sensor 3-wire/4-wire (Pt100: JIS C 1604-1997, JPt100: JIS C 1604-1989)
Accuracy
Pt100, JPt100: ±0.05% f.s. ±0.5ËC
Max. allowable input
60 V DC (maximum voltage between input terminals that does not cause
damage)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
Voltage Inputs
(accuracy is specified at 23 ±5ËC, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Input Terminals/Impedance Same as for thermocouples, Input resistance: 1 MΩ
Measurement
ranges
Accuracy
100 mV f.s. : -100 to +100 mV (5 μV resolution)
1V f.s. : -1 to +1 V (50 μV resolution)
10V f.s. : -10 to +10 V (500 μV resolution)
100V f.s. : -60 to +60 V (5 mV resolution)
1 to 5 V f.s. : 1 to 5 V (500 μV resolution)
±0.1 % f.s. (in the 1-5 V f.s. range only, f.s. = 10 V)
Max. allowable input
60 V DC (maximum voltage between input terminals that does not cause
damage)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
Humidity measurement
(only in 8420-51/8421-51, accuracy specified @23 ±5ËC, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Input terminals Commonly used with thermocouples
Measurement ranges 100% rh f.s. : 5.0 to 95.0% rh (0.1% rh resolution)
Accuracy
Refer to the accuracy table for the HUMIDITY SENSOR 9701, 9681 or 9653.
Filter function (common to thermocouple/platinum resistance temperature sensor/voltage/humidity inputs)
Digital Filter
OFF, 50 Hz, 60 Hz (With 50 and 60 Hz settings, the digital ï¬lter is
automatically set according to recording interval)
Pulse inputs (@23 ±5ËC, accuracy guaranteed for one year)
Input terminals
Custom connector
(up to 4 channel inputs using the CONNECTION CABLE 9641)
Input signal
condition
â No-voltage 'a' contact (normally open contact), open collector or
voltage input
â Insulated between all pulse input channels and instrument
chassis, and between each pulse input channel
â Maximum input voltage: 15 V DC
â Detection level: HIGH = at least 1.0 V, LOW = 0 to 0.5 V
â Pulse input cycle: with ï¬lter OFF, 200 μs or more (both H and L
periods must be at least 100 μs), with ï¬lter ON, 100 ms or more (both H
and L periods must be at least 50 ms)
â Slope: Rising or falling edge can be set for each channel
â Chatter-prevention ï¬lter: can be set on/off
50,000 c f.s. : 0 to 50,000 counts (1-count resolution)
Pulse totalization
ranges
500,000 c f.s. : 0 to 500,000 counts (10-count resolution)
5M c f.s. : 0 to 5M counts (100-count resolution)
100M c f.s. : 0 to 100M counts (2,000-count resolution)
2,500M c f.s. : 0 to 2,500M counts (50,000-count resolution)
Rotation range
5,000/n (r/s) f.s. : 0 to 5,000/n (r/s); Resolution: 1/n (r/s)
Note: n = pulses per rotation (1 to 1,000)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
â Use Model 9701 or 9681 only with Models 8420-51/8421-51, â Model 9653 can be used with Models 8420-51, 8421-51,
cable length: 3m, and if in combination with Models 9701, the
8420-01 or 8421-01.
main unit must be installed with firmware version 3.05 or later.
â Optional Accessories (sold separately)
PRINTER UNIT 8992 Specifications (installs on HiLOGGERs)
Recording paper
Recording width: 100mm (10 divisions full scale), the RECORDING
PAPER 9234 (112 mm (4.41 in) Ã 18 m (59.06 ft) roll, thermal paper )
Recording speed Maximum 2 mm/s (using AC Adapter or battery pack)
Functions
â Logging (prints numerical measurement values)
â Hybrid (prints waveforms simultaneously with numerical values for each
numerical quantity at selectable printing intervals)
â Real-time printing (automatic printing during measurement, available
when waveform time axis is 5 s to 1 d, and numeric printing interval is at least 5 s)
â Manual printing (while measuring, on/off selectable)
â Other functions include selectable printing between A/B
cursors, screen hard copy, list print, report print, list/gauge,
comment print
DIGITAL I/O UNIT 8993 Specifications (installs on HiLOGGERs)
Logic Inputs
â Screw-type terminal block, 16 channels (common GND between all input
channels to instrument chassis, and between each input)â Input signal condition:
No-voltage 'a' contact (normally open contact), open collector or voltage
input, â Detection level: HIGH = at least 2.5 V, LOW = 0 to 1.5 V,
â Maximum input voltage: 50 V DC, â Input resistance: 1.1 MΩ
Alarm outputs
â 16 isolated channels: from any of 32 analog input channels
and 4 pulse input channels, 16 digital input bits (= 1 channel) can
be output for any of 16 channels (all output channels isolated from the
chassis and from each other), â Output latch settings: latch/no latch,
â alarm sound: enabled/disabled, â Max. rated voltage to earth:
60 V DC, â Output type: open collector (active low), â Maximum
switching capability: 5 to 60 V DC @10 mA*5
*5 Mechanical contact outputs are available by modiï¬cation for users who want
to control sequencers using relays or photocouplers that require drive current
exceeding 10 mA. Please contact your distributor or HIOKI for details.
â Output refresh: at each recording interval
|
▷ |