■ Product Specifications
Functional Specifications
Display
5.7-inch STN Color LCD (240 × 320 dots). Displays either waveforms
only, waveforms and numerical values simultaneously, numerical
values only or calculation results and waveforms simultaneously.
Japanese/English language selectable. The most recent 16 MB of
data (with one-channel recording) can be displayed by back scrolling.
Waveform compression
and magnification
Time-axis: 1/2/5/10/20/30 seconds, 1/2/5/10/20/30 minutes,
1/2/5/10/12 hours, 1 day/division
Voltage-axis zoom: ×100, ×50, ×20, ×10, ×5, ×2, ×1, ×1/2
Search function Event marks can be searched
Calculation
functions
Waveform parameter calculations: Up to four simultaneous calculations
are supported: average, peak, maximum, minimum and effective values,
timing of maximum and minimum values, and waveform processing
calculation: displays sums and differences between channel waveforms
Communication
functions
(Controlled by PC)
■ Data can be collected in real time using the optional LOGGER
COMMUNICATOR 9334 (data collection software application).
■ Data in internal memory*1 and in the PC Card can be
transferred by the FTP server function.
*1 Data in internal memory can be transferred only when not measuring.
■ Remote control by communications commands (separate
programming required)
■ Remote control by HTTP server function (with no special software
other than a common Web browser, HiLOGGER settings, data acquisition and
screen monitoring can be performed)
Communication
functions
(Sending to PC)
■ Automatic data sending by FTP client function (periodically
during measurement or after measuring, a file automatically saved to PC Card is
automatically sent to an FTP server on the internal network or on a remote PC)
■ Automatic e-mail sending (upon trigger start/stop, trigger warning,
power outage recovery, memory or card full status, an e-mail notification is sent to
a PC on the local network or to a remote PC through an SMTP mail server)
Communication
functions (PPP)
To communicate with a modem-equipped PC through the public
telephone circuit, connect a standard modem to the RS-232C
terminal using the RS-232C CABLE 9721.
Miscellaneous
Waveform scroll, cursor measurement, scaling, automatic saving
(binary/text selectable), start condition retention, settings saving, comment
entry, event marking (for search), automatic setup, saves the most recent
16 MB of data (with one-channel recording) in internal memory
Trigger Functions
Trigger source Analog input: all channels, Pulse totalizer inputs: P1 to P4, Logic inputs
(conditions can be set for LI-1 to LI-16 (using the DIGITAL I/O UNIT 8993), External trigger, Time
each channel)
trigger, Logical Product of each trigger source (AND), Logical Sum (OR)
External trigger
Active low, valid pulse width H period 2.5 ms or more, L period
2.5 ms or more (with external trigger filter ON)
Trigger timing Start, Stop and Start/Stop (separate trigger conditions can be set to start and stop)
Trigger type
(analog, pulse)
Level: Triggers when rising or falling through preset level
Window: Triggers when entering or exiting range defined by
preset upper and lower limit values
Trigger level resolution 0.5% f.s. (f.s. = 10 divisions)
Trigger type (logic) Pattern trigger by 1, 0 and × (don't care) (with the DIGITAL I/O UNIT 8993)
Miscellaneous
Pre-trigger (records period before trigger, can be set for real-time saving),
Trigger output (open collector, 5V output, active low, at least 100 ms pulse
width), Trigger mode (single, continuous)
Thermocouple Inputs
(accuracy specified @23 ±5˚C, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Screw-type terminal block (recommended minimum wire diameter 0.32 mm*2),
Input Terminals/
removable, supplied terminal block cover (all channel terminals are isolated from
each other and chassis), Input impedance: 1MΩ (850kΩ when open-circuit polling is
Impedance
enabled)
*2 Recommended wire: 0.14 to 1.5 mm2 single strand, or 0.14 to 1.0 mm2 twisted multi strand.
To connect smaller thermocouple wire, use the TERMINAL UNIT 9329.
Setting range
*3 Upper and lower limits depend on the
100˚C f.s. : -100 to 100˚C*3 (0.01˚C resolution)
500˚C f.s. : -200 to 500˚C*3 (0.1˚C resolution)
measurement input range of each sensor 2000˚C f.s. : -200 to 2000˚C*3 (0.5˚C resolution)
Thermocouples
are JIS C 1602-1995 compliant
(except type W)
K: -200 to 1350˚C, E: -200 to 1000˚C, J: -200 to 1200˚C, T: -200 to 400˚C,
N: -200 to 1300˚C, R: 400 to 1700˚C, S: 400 to 1700˚C, B: 400 to 1800˚C,
W (WRe5-26): 400 to 2000˚C
Note: Types W, R and S can measure from 0˚C, but measurements in this range are not
guaranteed Standard. Reference contact: internally and externally switchable
Accuracy
Types K, E, J, T, N: ±0.05% f.s. ±1˚C
Types R, S, B, W: ±0.05% f.s. ±2˚C (applicable at 400˚C and above)
Standard reference contact accuracy: ±1˚C
(with internal compensation, add to measurement accuracy)
Max. allowable input
60 V DC (maximum voltage between input terminals that does not cause
damage)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
Platinum resistance temperature sensor inputs
(only in 8420-51/8421-51, accuracy specified @23 ±5˚C, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Input terminal, Common thermocouple application, detection current: 1 mA,
detection current Input Resistance: 2 MΩ
Setting range 100˚C f.s. : -100 to 100˚C*4 (0.01˚C resolution)
*4 Upper and lower limits depend
on the measurement input range
of each sensor
500˚C f.s. : -200 to 500˚C*4 (0.1˚C resolution)
2000˚C f.s. : -200 to 2000˚C*4 (0.5˚C resolution)
Resistance
Pt100: -200 to 800˚C, JPt100: -200 to 500˚C, Connection:
temperature sensor 3-wire/4-wire (Pt100: JIS C 1604-1997, JPt100: JIS C 1604-1989)
Accuracy
Pt100, JPt100: ±0.05% f.s. ±0.5˚C
Max. allowable input
60 V DC (maximum voltage between input terminals that does not cause
damage)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
Voltage Inputs
(accuracy is specified at 23 ±5˚C, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Input Terminals/Impedance Same as for thermocouples, Input resistance: 1 MΩ
Measurement
ranges
Accuracy
100 mV f.s. : -100 to +100 mV (5 μV resolution)
1V f.s. : -1 to +1 V (50 μV resolution)
10V f.s. : -10 to +10 V (500 μV resolution)
100V f.s. : -60 to +60 V (5 mV resolution)
1 to 5 V f.s. : 1 to 5 V (500 μV resolution)
±0.1 % f.s. (in the 1-5 V f.s. range only, f.s. = 10 V)
Max. allowable input
60 V DC (maximum voltage between input terminals that does not cause
damage)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
Humidity measurement
(only in 8420-51/8421-51, accuracy specified @23 ±5˚C, from 30 minutes after power on and after zero point adjustment, guaranteed for one year)
Input terminals Commonly used with thermocouples
Measurement ranges 100% rh f.s. : 5.0 to 95.0% rh (0.1% rh resolution)
Accuracy
Refer to the accuracy table for the HUMIDITY SENSOR 9701, 9681 or 9653.
Filter function (common to thermocouple/platinum resistance temperature sensor/voltage/humidity inputs)
Digital Filter
OFF, 50 Hz, 60 Hz (With 50 and 60 Hz settings, the digital filter is
automatically set according to recording interval)
Pulse inputs (@23 ±5˚C, accuracy guaranteed for one year)
Input terminals
Custom connector
(up to 4 channel inputs using the CONNECTION CABLE 9641)
Input signal
condition
■ No-voltage 'a' contact (normally open contact), open collector or
voltage input
■ Insulated between all pulse input channels and instrument
chassis, and between each pulse input channel
■ Maximum input voltage: 15 V DC
■ Detection level: HIGH = at least 1.0 V, LOW = 0 to 0.5 V
■ Pulse input cycle: with filter OFF, 200 μs or more (both H and L
periods must be at least 100 μs), with filter ON, 100 ms or more (both H
and L periods must be at least 50 ms)
■ Slope: Rising or falling edge can be set for each channel
■ Chatter-prevention filter: can be set on/off
50,000 c f.s. : 0 to 50,000 counts (1-count resolution)
Pulse totalization
ranges
500,000 c f.s. : 0 to 500,000 counts (10-count resolution)
5M c f.s. : 0 to 5M counts (100-count resolution)
100M c f.s. : 0 to 100M counts (2,000-count resolution)
2,500M c f.s. : 0 to 2,500M counts (50,000-count resolution)
Rotation range
5,000/n (r/s) f.s. : 0 to 5,000/n (r/s); Resolution: 1/n (r/s)
Note: n = pulses per rotation (1 to 1,000)
Max. rated voltage 60 V DC (Upper limit voltage that does not cause damage when applied
to earth
between input channel and chassis, and between each input channels)
■Use Model 9701 or 9681 only with Models 8420-51/8421-51, ■Model 9653 can be used with Models 8420-51, 8421-51,
cable length: 3m, and if in combination with Models 9701, the
8420-01 or 8421-01.
main unit must be installed with firmware version 3.05 or later.
■ Optional Accessories (sold separately)
PRINTER UNIT 8992 Specifications (installs on HiLOGGERs)
Recording paper
Recording width: 100mm (10 divisions full scale), the RECORDING
PAPER 9234 (112 mm (4.41 in) × 18 m (59.06 ft) roll, thermal paper )
Recording speed Maximum 2 mm/s (using AC Adapter or battery pack)
Functions
■ Logging (prints numerical measurement values)
■ Hybrid (prints waveforms simultaneously with numerical values for each
numerical quantity at selectable printing intervals)
■ Real-time printing (automatic printing during measurement, available
when waveform time axis is 5 s to 1 d, and numeric printing interval is at least 5 s)
■ Manual printing (while measuring, on/off selectable)
■ Other functions include selectable printing between A/B
cursors, screen hard copy, list print, report print, list/gauge,
comment print
DIGITAL I/O UNIT 8993 Specifications (installs on HiLOGGERs)
Logic Inputs
■ Screw-type terminal block, 16 channels (common GND between all input
channels to instrument chassis, and between each input)■ Input signal condition:
No-voltage 'a' contact (normally open contact), open collector or voltage
input, ■ Detection level: HIGH = at least 2.5 V, LOW = 0 to 1.5 V,
■ Maximum input voltage: 50 V DC, ■ Input resistance: 1.1 MΩ
Alarm outputs
■ 16 isolated channels: from any of 32 analog input channels
and 4 pulse input channels, 16 digital input bits (= 1 channel) can
be output for any of 16 channels (all output channels isolated from the
chassis and from each other), ■ Output latch settings: latch/no latch,
■ alarm sound: enabled/disabled, ■ Max. rated voltage to earth:
60 V DC, ■ Output type: open collector (active low), ■ Maximum
switching capability: 5 to 60 V DC @10 mA*5
*5 Mechanical contact outputs are available by modification for users who want
to control sequencers using relays or photocouplers that require drive current
exceeding 10 mA. Please contact your distributor or HIOKI for details.
■ Output refresh: at each recording interval