English
Language : 

DS-0097 Datasheet, PDF (4/4 Pages) All Sensors Corporation – MiniatureAmplifiedPressureSensors
Performance Characteristics for 5 cmH2O-D1-P4V-MINI
Parameter, note 1
Operating Range, pressure
Output Voltage, @5 cmH2O
Output Voltage @ zero pressure
Output Voltage @-0.2 cmH2O
Offset Temperature Shift (-25 to 85°C), note 2
Offset Warm-up Shift, note 3
Offset Position Sensitivity (±1g)
Offset Long Term Drift (one year)
Linearity, hysteresis error, note 4
Span Shift (-25 to 85°C), note 2
Minimum
-0.2
4.40
0.30
0.13
Nominal
4.5
0.35
0.18
±15
±15
±20
0.05
Maximum
5
4.6
0.40
0.23
±1.0
±0.25
±1.0
Performance Characteristics for 20 cmH2O-D1-P4V-MINI
Parameter, note 1
Operating Range, pressure
Output Voltage, @20 cmH2O
Output Voltage @ zero pressure
Output Voltage @-0.2 cmH2O
Offset Temperature Shift (-25 to 85°C), note 2
Offset Warm-up Shift, note 3
Offset Position Sensitivity (±1g)
Offset Long Term Drift (one year)
Linearity, hysteresis error, note 4
Span Shift (-25 to 85°C), note 2
Minimum
-0.2
4.40
0.30
0.25
Nominal
4.5
0.35
0.30
±10
±10
±10
0.05
Maximum
20
4.6
0.40
0.35
±1.0
±0.25
±1.0
Units
cmH2O
volt
volt
volt
%span
mvolt
mvolt
mvolt
%fs
%span
Units
cmH2O
volt
volt
volt
%span
mvolt
mvolt
mvolt
%fs
%span
Performance Characteristics for 120 cmH2O-D1-P4V-MINI
Parameter, note 1
Operating Range, pressure
Output Voltage, @120 cmH2O
Output Voltage @ zero pressure
Output Voltage @-10 cmH2O
Offset Temperature Shift (-25 to 85°C), note 2
Offset Warm-up Shift, note 3
Offset Position Sensitivity (±1g)
Offset Long Term Drift (one year)
Linearity, hysteresis error, note 4
Span Shift (-25 to 85°C), note 2
Minimum
Nominal
Maximum
Units
-10
120
cmH2O
4.40
4.5
4.6
volt
0.30
0.35
0.40
volt
0.23
0.28
0.33
volt
±1.0
%span
±15
mvolt
±15
mvolt
±20
mvolt
0.05
±0.25
%fs
±1.0
%span