DS-0097 Datasheet, PDF (2/4 Pages) All Sensors Corporation – MiniatureAmplifiedPressureSensors
Pressure Sensor Characteristics Maximum Ratings
Supply Supply Voltage VS
Common-mode pressure
Lead Temperature
(soldering 2-4 sec.)
+4.5 to
+75.5Vdc
10 psig
250°C
Standard Pressure Ranges
Environmental Specifications
Temperature Ranges
Compensated
Operating
Storage
Humidity Limits
see specification
-25 to 85° C
-40 to 125° C
0 to 95% RH
(non condensing)
Part Number
5 cmH2O-D1-4V-MINI
20 cmH2O-D1-4V-MINI
120 cmH2O-D1-4V-MINI
5 cmH2O-D1-P4V-MINI
20 cmH2O-D1-P4V-MINI
120 cmH2O-D1-P4V-MINI
Operating Pressure
-0.2 to 5 cmH2O
-0.2 to 20 cmH2O
-10 to 120 cm H2O
-0.2 to 5 cmH2O
-0.2 to 20 cmH2O
-10 to 120 cm H2O
Compensated Range
5 to 50° C
5 to 50° C
5 to 50° C
-25 to 85° C
-25 to 85° C
-25 to 85° C
Proof Pressure
50 cmH2O
100 cmH2O
300 cmH2O
50 cmH2O
100 cmH2O
300 cmH2O
Burst Pressure
200 cmH2O
300 cmH2O
600 cmH2O
200 cmH2O
300 cmH2O
600 cmH2O
Specification Notes
NOTE 1: ALL PARAMETERS ARE MEASURED AT 5.0 VOLT EXCITATION, FOR THE NOMINAL FULL SCALE PRESSURE AND ROOM TEMPERATURE UNLESS
OTHERWISE SPECIFIED. PRESSURE MEASUREMENTS ARE WITH POSITIVE PRESSURE APPLIED TO THE FRONT PORT.
NOTE 2: SHIFT IS RELATIVE TO 25°C.
NOTE 3: SHIFT IS WITHIN THE FIRST HOUR OF EXCITATION APPLIED TO THE DEVICE.
NOTE 4: MEASURED AT ONE-HALF FULL SCALE RATED PRESSURE USING BEST STRAIGHT LINE CURVE FIT.
NOTE 5: THE VOLTAGE ADDED TO THE OFFSET VOLTAGE AT FULL SCALE PRESSURE. NOMINALLY THE OUTPUT VOLTAGE RANGE IS 0.25 TO 4.25 VOLTS
FOR MINUS TO PLUS FULL SCALE PRESSURE.
All Sensors reserves the right to make changes to any products herein. All Sensors does not assume any liability arising out of the application or
use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
Equivalent Circuit
Miniature Amplified Medical Range Pressure Sensors