TLP150 Medical Series Data Sheet
Ordering Information
Output Voltage
Min
Output Currents
Max (free air)(1, 4) Max (forced air) (2, 4)
12 V
0A
24 V
0A
8.4 A
4.2 A
12.5 A
6.3 A
Ripple
120 mV
240 mV
Total Regulation
± 3.0%
± 3.0%
Model Numbers (8, 10)
TLP150N-99S12J
TLP150N-99S24J
Notes:
1. Free air convection. Maximum continuous output power not to exceed 100 W. Refer to Figure 1 for the derating curve.
2. 200 LFM forced air cooling from the ac input side. Maximum continuous output power not to exceed 150 W.
3. Figure is peak-to-peak for room temperature rating. Output noise measurements are made across a 20 MHz bandwidth using a 6 inch twisted pair, terminated with a 10 μF
tantalum capacitor and a 0.1 μF ceramic capacitor.
4. CAUTION: Allow a minimum of 1 second after disconnecting line power when making thermal measurements. For optimum reliability no part of the heatsink should exceed
115 °C and no semi-conductor case temperature should exceed 120 °C.
5. No external filtering required during conducted emissions testing but some applications may require additional filtering to achieve system compliance. Compliance with
radiated EMI specifications may require mounting in a suitable enclosure.
6. This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product.
7. Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G.
8. Replace the ‘J’ at the end of the model number with ‘FJ’ when the optional standby output and/or remote ON/OFF control is required e.g. TLP150N-99S12FJ.
9. 12 V (fan) present when main output is present. An optional 5 Vsb (standby) output is available whenever ac input is present, regardless of remote ON/OFF signal status.
10.The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant.
11.NOTICE: Some models do not support all options. Please contact your local Artesyn representative or use the on-line model number search tool at http://www.artesyn.com/
power to find a suitable alternative.
12.Power good signal required 100 mA load on the main output.
13.This product is a Component Power Supply and is only for inclusion by professional installers within other equipment and must not be operated as a standalone product. EMC
compliance to appropriate standards must be verified at the system level. This product is for sale to OEMs and System Integrators, including through Distribution Channels. It is
not intended for sale to End Users.
DERATING CURVE Output Power (Watts)
Outline Dimensions
5.000 (127.00)
150 W
100 W
200 LFM (1.0 m/s)
FORCED AIR COOLING
75 W
4.550 (115.57)
J3
0W
0C
NATURAL CONVECTION COOLING
10 C 20 C 30 C 40 C 50 C 60 C
Figure 1: Derating Curve
50 W
70 C
J1 1
Neutral
Live
J2
Ground/Earth
ø0.156 (4 Places)
Dimensions in Inches (mm)
1
FAN 1
J4
1
R88
0.225 (5.72)
Trimmer Adjust
Output Voltage
Figure 2: Mechanical Drawing
1 J301
Total Height
1.250 (31.75)