English
Language : 

LPT62-M Datasheet, PDF (1/3 Pages) Artesyn Technologies – Total Power: 60 - 80 Watts
POWER
Data Sheet
Total Power: 60 - 80 Watts
Input Voltage: 85 - 264 Vac
120 - 300 Vdc
# of Outputs: Triple
SPECIAL FEATURES
ƒƒ Medical safety approvals
ƒƒ Universal input
ƒƒ 3” x 5” footprint
ƒƒ Remote sense on main output
ƒƒ Built-in EMI filter
ƒƒ Low output ripple
ƒƒ Adjustable 5V output
ƒƒ Overvoltage protection
ƒƒ Overload protection
ƒƒ 110 KHz switching frequency
ƒƒ LPX60 enclosure kit available
ƒƒ RoHS compliant
SAFETY
ƒƒ VDE
ƒƒ UL
ƒƒ CSA
ƒƒ CE
0750/EN60601-1 (IEC601)
21310-3336-0015
UL2601
E182560
CSA 22.2 No. 601.1
LR109492
Mark (LVD)
LPT60-M Series
60 Watts
Electrical Specifications
Input
Input range
85 - 264 Vac
120 - 300 Vdc
Frequency
47 - 440 Hz
Inrush current
<18 A peak @ 115 Vac;
<36 A peak @ 230 Vac, cold start @ 25 °C
Input current
1.5A max. (RMS) @ 115VAC
Efficiency
70% typical at full load
EMI
FCC Class A conducted; CISPR 22 Class A conducted EN55022 Class
A conducted; VDE 0878 PT3 Class A conducted
Safety ground
leakage current
<75 μA max. @ 50/60 Hz, 264 Vac input
Output
Maximum power
60 W for convection; 80 W with 30CFW forced air
Cross regulation
±2% on output 1; ±5% on outputs 2, 3
Adjustment range
-5, +10% minimum
Hold-up time
20 ms @ 60 W load, 115 Vac nominal line
Overload protection
Short circuit protection on all outputs. Case overload protected @ 110%
to 145% above peak rating
Overvoltage protection 5.7 to 6.7 Vdc on the main output
Remote sense
Compensates for 0.5 V lead drop min. Will operate without remote
sense connected. Reverse connection protected.