Specifications
Environmental
Body: -40°C to 60°C, 0 to 95%RH non condensing
• External probe: -196°C to 200°C
• Logger Dimensions: 190 mm x 127 mm x 45 mm
(7.5 in x 5 in x 1.75)
• Base Station Dimensions: 6 4 mm x 32 mm x 15 mm
(2.5 in x 1.25 in x 5 in)
Temperature
Single Channel External Sensor
• ±0.1°C, range 0°C to +60°C
• ±0.2°C, range -22°C to 0°C
• ±0.5°C, range -80°C to –22°C
• ±2.0°C, @ -196°C
For 2 and 5 Channel Temperature Loggers:
±0.1C range –80°C to 130°C
RH
• 2% from 10 to 90%RH
• 5% from 0 to 10%, 90 to 100%RH
• 3% at 40°C, 75%RH
• Certificate states ICH Points
Auxiliary Inputs
• Contact: dry
• Voltage and current 0.5% FS
Battery Life (Varies ±10%)
1 min sample rate: 8000 hours
System Documentation
IQ/OQ Protocol
The Installation Qualification/Operational Qualification
Protocol defines a set of procedures to ensure that
the Kaye Validator system is properly installed and
operated correctly, and is adequately documented
and controlled according to cGMP requirements. The
documents are provided in hard copy and on CD,
allowing users to modify the documentation to suit
specific organizational requirements.
Updates and addenda are available through the Kaye
website.
Services
We provide a complete aray of services for the RF
ValProbe® product range:
• Onsite installation,commissioning and IQ/OQ for
RF ValProbe® system.
• Operator training and technical support.
• Maintenance agreements.
• Factory calibration.
• Rental services.
Validation Reference
The Kaye Validator system is supported with Design
Qualification documentation. The Validation Reference
Binder provides a comprehensive overview of our
Quality Policy, description of ISO 9001 implementation
and support procedures, and standards for the
development, testing, and maintenance of hardware
and software. Quality Control documents, Development
procedures, Quality Assurance procedures, Release
documents, and Quality Assurance test documentation
are all included.