English
Language : 

STBHL02_15 Datasheet, PDF (3/11 Pages) Analog Devices – Analog Signal Conditioning Panels
5B02, STB-TCI, STB-HLI, STB-TC, STB-HL02, STB-AOT
STB-TCI–SPECIFICATIONS STB-TCI
The STB-TCI isolated thermocouple panel provides the means
to interface up to eight low level voltage signals requiring high
electrical isolation. Low level signals, including thermocouple
(typical @ +25؇C and +5 V dc)
voltages, can be read with high accuracy even in noisy environ-
Number of Input Channels
8
ments. Standard thermocouple types such as J, K, T, E, R, B
Isolation
750 V rms Channel-to-Channel
and S can be used with the panel, and cold junction compensation
and Channel-to-System
is provided. Each input channel is provided with user selectable
gains of 50, 100, 200 and 1000, permitting signal ranges of
± 5 mV, ± 25 mV, ± 50 mV and ± 100 mV. Each channel has a
quick-disconnect screw terminal connector making field installa-
tion and replacement easy. A cover (CV-01) and a rack mount
(RM-02) may be ordered separately.
Input Impedance
Input Voltage Range
Input Overvoltage Protection
Gain Range
50 MΩ
± 5 mV, ± 25 mV, ± 50 mV,
± 100 mV
240 V ac rms
50, 100, 200, 1000
OBSOLETE Each input signal on the STB-TCI panel passes through a 3-pole
low-pass filter with a cutoff frequency of 4 Hz. This filter re-
duces noise that may be present at the analog inputs. Break
detect circuitry is also present to determine if a thermocouple or
cable has been disconnected. Each channel has an isolation
amplifier that provides 750 V rms of channel-to-system and
channel-to-channel isolation. Overvoltage input protection of
240 V is provided as well. A functional diagram of the STB-TCI
is shown in Figure 2.
The STB-TCI, or the 5B02 with 5B30, 5B37, 5B40, or 5B47
modules, should be used instead of the STB-TC panel when high
isolation is required for a harsh environment. The STB-TCI panel
may be used instead of 5B Series modules when an application
has many low level inputs and single channel modularity is not
required.
Gain Accuracy1
Output Offset1
Offset Drift vs. Temperature
(Referred to Input)
CJC Accuracy
Low-Pass Filter Frequency
Common-Mode Rejection
Power Requirements
Physical Dimensions
Without Optional Cover
With Optional Cover
Jumper Configurable per
Channel
± 0.05% (at Factory Adjusted
Gain of 100)
± 3 mV
± 3.25 µV/°C (at Gain of 100)
± 1°C
4 Hz
130 dB
+5 V ± 5% @ 300 mA
8.25" × 6.8" × 1.97"
(20.95 cm × 17.27 cm × 5.0 cm)
8.25" × 6.8" × 2.14"
(20.95 cm × 17.27 cm × 5.43 cm)
Operating Temperature Range 0°C to +60°C
NOTES
1Adjustable to zero.
Specifications subject to change without notice.
REV. A
USER-SELECTABLE GAIN OF
50, 100, 200 OR 1000
CHANNEL 0
ANALOG SIGNAL
9
TO
TO
RTI-820
1
USER SELECTABLE GAIN
CHANNEL 1
ETC.
FROM
RTI-820
BOARD AND
CHANNEL ADDRESS
DECODING
COLD
JUNCTION
COMPENSATION
SENSOR
Figure 2. STB-TCI Block Diagram
CHANNEL 7
–3–