English
Language : 

LPS102-M Datasheet, PDF (1/3 Pages) Artesyn Technologies – Total Power: 80 - 150 Watts
POWER
Data Sheet
Total Power: 80 - 150 Watts
Input Voltage: 90 - 264 Vac
120 - 300 Vdc
# of Outputs: Single
SPECIAL FEATURES
ƒƒ Medical and ITE safety approvals,
2X MOPP
ƒƒ Active power factor correction
ƒƒ 2” x 4” footprint
ƒƒ Less than 1U high
ƒƒ EN61000-3-2 compliant
ƒƒ Remote sense
ƒƒ Power fail
ƒƒ Adjustable main output
ƒƒ Built-in Class B EMI filter
ƒƒ Overvoltage protection
ƒƒ Overload protection
ƒƒ Thermal overload protection
ƒƒ Isolated 12 V fan output
ƒƒ LPX100 Enclosure kit available
ƒƒ POE isolation on main output of
LPS109-M
SAFETY
ƒƒ TUV
ƒƒ UL
ƒƒ cULus
ƒƒ CB
ƒƒ CE
ƒƒ CQC
60950, 60601-1
60950, 60601-1
60950, 60601-1
Certificate & report
Mark (LVD)
Mark
LPS100-M Series
150 Watts
Electrical Specifications
Input
Input ra­nge
90 - 264 Vac; 120 - 300 Vdc
Frequency
47 - 63 Hz
Inrush current
50 A max., cold start @ 25 °C
Efficiency
88% typical at full load
EMI/RFI
FCC Class B conducted; CISPR22 Class B conducted;
EN55022 Class B conducted; VDE0878PT3 Class B conducted
Safety ground leakage 275 μA @ 50/60 Hz, 264 Vac input
current
Output
Maximum power
100 W for convection (80 W for LPS102-M);
150 W with 30 CFM forced air (120 W for LPS102-M)
Adjustment range
± 10% minimum on the main outputs
Fan output
12 V @ 1 A isolated, ± 10%
Hold-up time
10 ms @ 150 W load, 120 Vac input
Overload protection
Short circuit protection on all outputs. Case overload
protected @ 110-160% above rating
Overvoltage protection 15-35% above nominal output
Logical Control
Power failure
Open collector logic signal goes high 100-500 msec after main output;
it goes low at least 6 msec before loss of regulation
Remote sense
Compensates for 0.5 V lead drop min. Will operate without remote
sense connected. Reverse connection protected.