English
Language : 

AHM85 Datasheet, PDF (30/40 Pages) XP Power Limited – IP22 Environmental Rating
XEnxxvxirxoxnxmxexnxtal
Characteristic
Operating Temperature
Case Temperature
(IEC60601 3rd Edition)
Storage Temperature
Cooling
Humidity
Operating Altitude
Ingress Protection
Shock
Vibration
Derating Curve
Figure 10
Minimum
0
-40
5
IP22
Typical
Maximum
+60
71
60
48
+85
95
3000
100%
60%
xppower.com
Units
ºC
ºC
ºC
%RH
m
Notes & Conditions
Derate linearly to 60% load at 60 ºC from
+40 °C. (See fig.10)
100% Load, with TAMB +40 ºC
60% Load Maximum, with TAMB +40 ºC
5% Load Maximum, with TAMB +40 ºC
Convection cooled, see fig.10
Non-condensing
3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Three axis 5-500 Hz at 2 g x 10 sweeps
0
10
20
40
60
Ambient Temperature (ºC)
Electromagnetic Compatibility - Immunity
Phenomenon
Low Voltage PSU EMC
Harmonic Current
ESD
Radiated
EFT
Surges
Conducted
Magnetic Field
Dips and Interruptions
Standard
EN61204-3
EN61000-3-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
EN60601-1-2
Test Level
High severity level
Class A
3
3
3
Installation class 3
3
3
Dip: 30% 500 ms
Dip: 60% 200 ms
Dip: 80% 5000 ms
Dip: 100% 5000 ms
Dip: 30% 25 AC Cycles
Dip: 60% 5 AC Cycles
Dip: 100% 0.5 AC Cycles
Int.: >95% 5000 ms
Criteria
as below
Notes & Conditions
A
A
A
A
A
A
A
B
A
B
A
230 VAC 100% load, 100 VAC 80% load
A
230 VAC 100% load, 100 VAC 80% load
A
B
6