SDU Datasheet, PDF (1/2 Pages) XP Power Limited – Medical Approvals
150 Watts
SDU Series
xppower.com
• Medical Approvals
• 3.2” x 5” Footprint
• Fits 1U Applications
• High Efficiency
• 12 V & 24 V Outputs
• U Channel or Open Frame Versions
• 3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
No Load Input Power
Earth Leakage Current
Input Protection
• 90-264 VAC
• 47-63 Hz
• 2.0 A at 90 VAC, 1 A at 230 VAC
• Max 100 A at 230 VAC, cold start 25 °C
• >0.9 at >40% load
• <1 W
• <250 µA at 264 VAC
• F5 A / 350 V internal in line fuse
Output
Output Voltage
• See table
Output Voltage Trim • ±10%
Initial Set Accuracy • ±1%
Minimum Load
• No minimum load required
Start Up Delay
• 3.0 s max at 120 VAC
Start Up Rise Time
• 30 ms typical
Hold Up Time
• 20 ms min at full load, 120 VAC
Line Regulation
• ±0.5%
Load Regulation
• ±2%
Over/Undershoot
• 3% max
Transient Response
• 4% max deviation, recovery to within
1% in 500 µs for a 50% load change
Ripple & Noise
• 1% pk-pk, 20 MHz bandwidth
Overvoltage Protection • 112-130% Vnom with auto recovery
Overload Protection • 110-185%
Short Circuit Protection • Trip & restart (hiccup mode), auto recovery
Temperature
Coefficient
• 0.04% /°C
General
Efficiency
Isolation
Protection Level
Switching Frequency
Power Density
MTBF
• Typically 87%
• 4000 VAC Input to Output
1500 VAC Input to Ground
1000 VDC Output to Ground
• Primary to secondary: 2 x MOPP
Primary to earth: 1 x MOPP
• 65 kHz PFC typical,
55 kHz - 65 kHz PWM typical at full load
• 8.4 W/in3
• 110 kHrs to MIL-HDBK-217F at 25 °C, GB
Environmental
Operating Temperature • 0 °C to +70 °C, derate at 2.5%/ °C from
+50 °C to +70 °C
Storage Temperature • -20 °C to +85 °C
Operating Humidity • 5-90%, non-condensing
Storage Humidity
• 5-95%, non-condensing
Cooling
• Convection-cooled
Operating Altitude
• 3000 m
Vibration
• 5-50 Hz, acceleration 7.35 m/s2 on
X, Y and Z axis
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Dips & Interruptions
Safety Approvals
• EN55011 Level B conducted & radiated
• EN61000-3-2 class A
EN61000-3-2 class C for loads ≥40%
• EN61000-4-2, level 3 Perf Criteria A
• EN61000-3-3
• EN61000-4-3, 3 V/m Perf Criteria A
• EN61000-4-4, level 2 Perf Criteria A
• EN61000-4-5, installation class 3,
Perf Criteria A
• EN61000-4-6, 3V Perf Criteria A
• EN61000-4-11, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms
Perf Criteria A, A, A, B high line,
A, B, A, B low line. See note 5
• ANSI AMMI ES60601-1:2005, CSA C22-2
No 60601-1:08, EN60601-1:2006