PCM50US2V5 Datasheet, PDF (1/4 Pages) XP Power Limited – AC-DC Power Supplies | |||
|
50/80/120 Watts
PCM Series
xppower.com
⢠4000 VAC Isloation
⢠Low Leakage Current
⢠Single, Dual and Triple Outputs
⢠Worldwide Medical Approvals
⢠CCC Approved (PCM80)
⢠Both Line and Neutral Fused
⢠Non-standard Connectors Available
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
⢠90-264 VAC
⢠47-63 Hz
⢠PCM50: 1.35 A max
PCM80: 1.30 A max
PCM120: 1.60 A max
⢠30 A at 115 VAC or
60 A at 230 VAC at 25 ËC cold start
⢠PCM50: 0.55 typical
PCM80/120: 0.94 min
⢠180 µA max at 132 VAC/60Hz
300 µA max at 264 VAC/50Hz
Output
Output Voltage
⢠See tables
Minimum Load
⢠No minimum load required for single
output models. See tables for multi
output models
Start Up Delay
⢠2 s max
Hold Up Time
⢠10 ms min at low line
Line Regulation
⢠±1% max at full load for ±10% change
in input
Load Regulation
⢠±5% max
Transient Response
⢠5% max. deviation, 500 µs recovery time
for a 50% load change
Ripple & Noise
⢠2% pk-pk, 20MHz BW
Overvoltage Protection ⢠112-132% with auto recovery on V1
Overcurrent Protection ⢠110-200% with auto recovery
on all outputs
Temperature
Coefficient
⢠±0.05% /ËC
General
Efficiency
Isolation
Switching Frequency
MTBF
⢠Single output models: 80% typical
Multi output models: 75% typical
⢠4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
⢠PCM50: 64 kHz typical
PCM80: 55 kHz typical
PCM120: 120 kHz typical
⢠100 kHrs at full load and 25 ËC ambient
per MIL-HDBK-217F
Environmental
Operating Temperature ⢠PCM50: 0 ËC to +70 ËC, derate from 100%
load at 50 ËC to 50% load at +70 ËC
PCM80: 0 ËC to +70 ËC, derate from 100%
load at 40 ËC to 50% load at +70 ËC
PCM120: 0 ËC to +60 ËC, derate from
100% load at 40 ËC to 50% load at +60 ËC
Storage Temperature ⢠PCM50/80: -40 ËC to +80 ËC
PCM120: -20 ËC to +85 ËC
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Power Frequency
Magnetic Fields
Dips & Interruptions
Safety Approvals
⢠EN55022, level B conducted
EN55022, level B radiated
⢠EN61000-3-2, class A
⢠EN61000-3-3 amendments 1 & 2
⢠IEC61000-4-2, level 3
Perf Criteria B
⢠IEC61000-4-3 10 V/m, level 3
Perf Criteria B
⢠IEC61000-4-4, level 3
Perf Criteria B
⢠IEC61000-4-5, level 3
Perf Criteria B
⢠IEC61000-4-6 10 V/m, level 3
Perf Criteria B
⢠IEC61000-4-8 10 A/m, Perf Criteria B
⢠IEC61000-4-11, 30% 500 ms,
60% 100 ms, 100 % 5000 ms
Perf Criteria A, A, C
⢠UL60601-1, CSA C22.2 No. 601.1 per cUL
EN60601-1, CCC (PCM80)
|
▷ |