PBM65US05 Datasheet, PDF (1/2 Pages) XP Power Limited – Medical Open Frame | |||
|
Medical Open Frame
65 Watts
PBM65 Series
THE XPERTS IN POWER
â¢
4000 VAC Isolation
â¢
Low Leakage Current
â¢
Meets Class B, EN55011 Conducted
â¢
Medical Safety Agency Approvals
â¢
Both Line and Neutral Fused
â¢
Fixed Switching Frequency
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Earth Leakage
Current
⢠85-264 VAC
⢠47-63 Hz
⢠1.6 A max at 115 VAC
1.0 A max at 230 VAC
⢠17 A max at 115 VAC
40 A max at 230 VAC
⢠90 µA max at 115 VAC, 60 Hz
150 µA max at 230 VAC, 50 Hz
Output
Output Voltage
⢠5.1 to 28 VDC
Output Voltage
Adjustment
⢠±5%
Output Power
⢠65 Watts
Minimum Load
⢠No minimum load required
Line Regulation
⢠±0.5% max at full load
Load Regulation
⢠±2% typical
Ripple and Noise
⢠±1% pk-pk max (See Note 2)
Transient Response ⢠4% maximum deviation, 500 µs
recovery time to within 1% for a
25% load change
Temperature
Coefficient
⢠±0.04%/°C
Hold Up Time
⢠10 ms minimum at 110 VAC
Overvoltage
Protection
Overcurrent
Protection
⢠112-132% with auto recovery
⢠125-200% with auto recovery
Short Circuit
Protection
⢠Trip & restart with auto recovery
Remote Sense
⢠Compensates for up to 0.5 V drop
General
Efficiency
â¢
MTBF
â¢
Switching Frequency â¢
Isolation Voltage
â¢
Size
â¢
Weight
â¢
Environmental
Operating
â¢
Temperature
Cooling
â¢
Relative Humidity
â¢
Storage
â¢
Temperature
EMC & Safety
Safety Approvals
â¢
EMI/EMC
â¢
ESD
â¢
Radiated Immunity â¢
EFT/Burst
â¢
Surge
â¢
66% minimum for 5 V models
75% minimum for all others
400,000 hrs min MIL-HDBK-217F
42 kHz ±5 kHz
4000 VAC Input to Output
1500 VAC Input to Output
500 VAC Output to Ground
5.0 x 3.0 x 1.2â
127.00 x 76.20 x 30.48 mm
330 g approx (PCB format)
0 °C to +70 °C
Derate linearly from 100% load at
+50 °C to 50% load at +70 °C
Convection cooled
5-95% non-condensing
-40 °C to +85 °C
UL2601-1, CSA C22.2 No 601.1,
EN60601-1 CE Mark (LVD)
EN55011 (CISPR11) Class B and
FCC Part 18 Level B radiated &
conducted
EN61000-4-2, Level 3 air,
Level 2 contact, Perf Criteria A
EN61000-4-3, 3 V/m, Perf Criteria A
EN61000-4-4, Level 3, Perf Criteria A
EN61000-4-5, Level 3, Perf Criteria A
â¢
www.xpplc.com
â¢
sales@xpplc.com
|
▷ |