English
Language : 

PBM65US05 Datasheet, PDF (1/2 Pages) XP Power Limited – Medical Open Frame
Medical Open Frame
65 Watts
PBM65 Series
THE XPERTS IN POWER
•
4000 VAC Isolation
•
Low Leakage Current
•
Meets Class B, EN55011 Conducted
•
Medical Safety Agency Approvals
•
Both Line and Neutral Fused
•
Fixed Switching Frequency
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Earth Leakage
Current
• 85-264 VAC
• 47-63 Hz
• 1.6 A max at 115 VAC
1.0 A max at 230 VAC
• 17 A max at 115 VAC
40 A max at 230 VAC
• 90 µA max at 115 VAC, 60 Hz
150 µA max at 230 VAC, 50 Hz
Output
Output Voltage
• 5.1 to 28 VDC
Output Voltage
Adjustment
• ±5%
Output Power
• 65 Watts
Minimum Load
• No minimum load required
Line Regulation
• ±0.5% max at full load
Load Regulation
• ±2% typical
Ripple and Noise
• ±1% pk-pk max (See Note 2)
Transient Response • 4% maximum deviation, 500 µs
recovery time to within 1% for a
25% load change
Temperature
Coefficient
• ±0.04%/°C
Hold Up Time
• 10 ms minimum at 110 VAC
Overvoltage
Protection
Overcurrent
Protection
• 112-132% with auto recovery
• 125-200% with auto recovery
Short Circuit
Protection
• Trip & restart with auto recovery
Remote Sense
• Compensates for up to 0.5 V drop
General
Efficiency
•
MTBF
•
Switching Frequency •
Isolation Voltage
•
Size
•
Weight
•
Environmental
Operating
•
Temperature
Cooling
•
Relative Humidity
•
Storage
•
Temperature
EMC & Safety
Safety Approvals
•
EMI/EMC
•
ESD
•
Radiated Immunity •
EFT/Burst
•
Surge
•
66% minimum for 5 V models
75% minimum for all others
400,000 hrs min MIL-HDBK-217F
42 kHz ±5 kHz
4000 VAC Input to Output
1500 VAC Input to Output
500 VAC Output to Ground
5.0 x 3.0 x 1.2”
127.00 x 76.20 x 30.48 mm
330 g approx (PCB format)
0 °C to +70 °C
Derate linearly from 100% load at
+50 °C to 50% load at +70 °C
Convection cooled
5-95% non-condensing
-40 °C to +85 °C
UL2601-1, CSA C22.2 No 601.1,
EN60601-1 CE Mark (LVD)
EN55011 (CISPR11) Class B and
FCC Part 18 Level B radiated &
conducted
EN61000-4-2, Level 3 air,
Level 2 contact, Perf Criteria A
EN61000-4-3, 3 V/m, Perf Criteria A
EN61000-4-4, Level 3, Perf Criteria A
EN61000-4-5, Level 3, Perf Criteria A
•
www.xpplc.com
•
sales@xpplc.com