PBM25US05 Datasheet, PDF (1/2 Pages) XP Power Limited – AC-DC Medical | |||
|
AC-DC Medical
25 Watts
PBM25 Series
THE XPERTS IN POWER
â¢
Small Size 2.50â x 4.25â x 0.96â
â¢
Low Leakage Current
â¢
Medical Safety Agency Approvals
â¢
Both Line and Neutral Fused
â¢
Meets Class B, EN55011 Conducted
â¢
Fixed Switching Frequency
Specification
Input
Input Voltage
â¢
Input Frequency
â¢
Input Current
â¢
Inrush Current
â¢
Earth
â¢
Leakage Current
Output
Output Voltage
â¢
Output Power
â¢
Minimum Load
â¢
Start Up Delay
â¢
Start Up Rise TIme â¢
Hold up Time
â¢
Initial Set Accuracy â¢
Line Regulation
â¢
Total Regulation
â¢
Transient Response â¢
Ripple & Noise
â¢
Overvoltage
â¢
Protection
Overcurrent
â¢
Protection
Short Circuit
â¢
Protection
Temperature
â¢
Coefficient
85-264 VAC
47-63 Hz
0.7 A (rms) at 115 VAC,
0.4 A (rms) at 230 VAC
10 A max at 115 VAC,
25 A max at 230 VAC
90 µA max at 115 VAC 60 Hz
150 µA max at 230 VAC 50 Hz
5.1 to 28 VDC, See Table
25/28 Watts, See Table
No minimum for single outputs models
See Table for multi outputs models
3 s max
12 ms
10 ms minimum at 110 VAC
±1%
±0.5% max at full load
See Table
4% max deviation, 500 µs recovery
time for a 25% load change
±1% max pk-pk (See Note 1)
Provided on output 1 only; set at
112-132% of its nominal output
voltage
125-200% with auto recovery
Trip & restart with auto recovery
±0.04%/°C
General
Efficiency
â¢
Isolation
â¢
Switching Frequency â¢
MTBF
â¢
Size
â¢
Weight
â¢
68% minimum for 5 V & multi output
models, 70% minimum all others
4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
42 KHz ±5 KHz
500,000 hrs per min MIL-HDBK-217F
4.25 x 2.50 x 0.96â
107.95 x 63.50 x 24.38 mm
Single output models 163 g,
Multi output models 175 g
Environmental
Operating
â¢
Temperature
Storage Temperature â¢
Cooling
â¢
Relative Humidity â¢
Operating Altitude â¢
Shock
â¢
Vibration
â¢
0 °C to +70 °C,
derate linearly from 100% load at
+50°C to 50% load at +70 °C
-40 °C to +85 °C
Convection cooled
5-95%, non-condensing
3000m
20 G in 3 directions, 3 times each
0.15mm over 10-55Hz and 1
octave/min. 3 axes for 30 min
EMC & Safety
EMC
â¢
ESD
â¢
Radiated Immunity â¢
EFT/Burst
â¢
Surge
â¢
Safety Approvals
â¢
EN55011 (CISPR11) Class B & FCC
Class B conducted
EN61000-4-2 Level 2 Perf Criteria A
EN61000-4-3 3 V/m Perf Criteria A
EN61000-4-4 Level 3 Perf Criteria A
EN61000-4-5 Level 3 Perf Criteria A
UL2601-1, CSA C22.2 No 601.1,
EN60601-1, IEC 601-1
â¢
â¢
www.xp-medical.com
sales@xp-medical.com
|
▷ |