English
Language : 

PBM25US05 Datasheet, PDF (1/2 Pages) XP Power Limited – AC-DC Medical
AC-DC Medical
25 Watts
PBM25 Series
THE XPERTS IN POWER
•
Small Size 2.50” x 4.25” x 0.96”
•
Low Leakage Current
•
Medical Safety Agency Approvals
•
Both Line and Neutral Fused
•
Meets Class B, EN55011 Conducted
•
Fixed Switching Frequency
Specification
Input
Input Voltage
•
Input Frequency
•
Input Current
•
Inrush Current
•
Earth
•
Leakage Current
Output
Output Voltage
•
Output Power
•
Minimum Load
•
Start Up Delay
•
Start Up Rise TIme •
Hold up Time
•
Initial Set Accuracy •
Line Regulation
•
Total Regulation
•
Transient Response •
Ripple & Noise
•
Overvoltage
•
Protection
Overcurrent
•
Protection
Short Circuit
•
Protection
Temperature
•
Coefficient
85-264 VAC
47-63 Hz
0.7 A (rms) at 115 VAC,
0.4 A (rms) at 230 VAC
10 A max at 115 VAC,
25 A max at 230 VAC
90 µA max at 115 VAC 60 Hz
150 µA max at 230 VAC 50 Hz
5.1 to 28 VDC, See Table
25/28 Watts, See Table
No minimum for single outputs models
See Table for multi outputs models
3 s max
12 ms
10 ms minimum at 110 VAC
±1%
±0.5% max at full load
See Table
4% max deviation, 500 µs recovery
time for a 25% load change
±1% max pk-pk (See Note 1)
Provided on output 1 only; set at
112-132% of its nominal output
voltage
125-200% with auto recovery
Trip & restart with auto recovery
±0.04%/°C
General
Efficiency
•
Isolation
•
Switching Frequency •
MTBF
•
Size
•
Weight
•
68% minimum for 5 V & multi output
models, 70% minimum all others
4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
42 KHz ±5 KHz
500,000 hrs per min MIL-HDBK-217F
4.25 x 2.50 x 0.96”
107.95 x 63.50 x 24.38 mm
Single output models 163 g,
Multi output models 175 g
Environmental
Operating
•
Temperature
Storage Temperature •
Cooling
•
Relative Humidity •
Operating Altitude •
Shock
•
Vibration
•
0 °C to +70 °C,
derate linearly from 100% load at
+50°C to 50% load at +70 °C
-40 °C to +85 °C
Convection cooled
5-95%, non-condensing
3000m
20 G in 3 directions, 3 times each
0.15mm over 10-55Hz and 1
octave/min. 3 axes for 30 min
EMC & Safety
EMC
•
ESD
•
Radiated Immunity •
EFT/Burst
•
Surge
•
Safety Approvals
•
EN55011 (CISPR11) Class B & FCC
Class B conducted
EN61000-4-2 Level 2 Perf Criteria A
EN61000-4-3 3 V/m Perf Criteria A
EN61000-4-4 Level 3 Perf Criteria A
EN61000-4-5 Level 3 Perf Criteria A
UL2601-1, CSA C22.2 No 601.1,
EN60601-1, IEC 601-1
•
•
www.xp-medical.com
sales@xp-medical.com