JHM1024D15 Datasheet, PDF (1/2 Pages) XP Power Limited – International Medical Approvals | |||
|
10 Watts
JHM10 Series
xppower.com
⢠International Medical Approvals
⢠4000 VAC Reinforced Insulation
⢠Meets IEC60601 1, 3rd Edition
⢠2 µA Patient Leakage Current
⢠DIP 24 Package
⢠EN55011 Level A With No External Components
⢠3 Year Warranty
Specification
Input
Input Voltage Range
Input Current
Inrush Current
Input Filter
Patient Leakage
Current
Undervoltage Lockout
Input Surge
⢠5 V (4.5-9 VDC)
12 V (9-18 VDC)
24 V (18-36 VDC)
⢠See table
⢠20 A max at 36 V
⢠Pi network
⢠2 µA max
⢠5 V models, on at >4.4 V, off <4.2 V
12 V models, on at >8.8 V, off <8.3 V
24 V models, on at >17.5 V, off <17.0 V
⢠5 V models 15 V for 3 s
12 V models 25 V for 3 s
24 V models 50 V for 3 s
Output
Output Voltage
⢠See table
Output Voltage Trim ⢠±10%
Minimum Load
⢠No minimum load required
Initial Set Accuracy ⢠±1% max on V1, ±2% max on V2
Start Up Delay
⢠5 ms typical
Start Up Rise Time
⢠2 ms typical
Line Regulation
⢠±0.3% max
Load Regulation
⢠±2% max 0% to 10% load,
±1% max 10% to 100% load
Cross Regulation
⢠±4% max on dual with one output set to
50% load and the other varied from
10% to 100% load (D05 is 20% to 100%)
Transient Response
⢠4% max deviation, recovery to within 1%
in <500 µs for a 50% load change at
0.25 A/µs rate
Ripple & Noise
⢠1% pk-pk max at 20 MHz bandwidth
Overload Protection ⢠120% - 200%, trip and restart
Overvoltage Protection ⢠115% - 140%
Temperature
Coefficient
⢠±0.03/°C max
Short Circuit Protection ⢠Trip and Restart (hiccup mode),
auto recovery
General
Efficiency
Isolation
Input to Output
Capacitance
⢠See tables
⢠4000 VAC for 1 min. double/reinforced with
a working voltage of 250 VAC. Meets 2 x
MOPP per 3rd edition of IEC60601-1
5000 VAC for 10 ms in accordance with
IEC60664-1
⢠20 pF max
Switching Frequency
Power Density
MTBF
⢠80 kHz to 1.2 MHz variable
⢠20.0 W/in3
⢠>1 MHrs typical to MIL-STD-217F
at 25 °C, GB
Environmental
Operating Temperature ⢠-40 °C to +80 °C, see derating curve
Case Temperature
⢠+100 °C max
Storage Temperature ⢠-40 °C to +100 °C
Operating Humidity ⢠5-90%, non-condensing
Cooling
⢠Natural convection
EMC & Safety
Emissions
Immunity
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Magnetic Field
Safety Approvals
⢠EN55011 & EN55022 level A conducted &
radiated with no external components
⢠IEC60601-1-2, EN61204-3
⢠EN61000-4-2, level 2, Perf Criteria A
⢠EN61000-4-3, 10 V/m Perf Criteria A
⢠EN61000-4-4, level 2 Perf Criteria A
⢠EN61000-4-5, level 1 Perf Criteria A
⢠EN61000-4-6, 10 Vm, Perf Criteria A
⢠EN61000-4-8, 3 A/m Perf Criteria A
⢠ANSI/AMMI ES60601-1 3rd Edition
CSA-22.2 No.60601-1:2008
IEC60601-1 3rd Edition
|
▷ |