JHM0612S05 Datasheet, PDF (1/2 Pages) XP Power Limited – International Medical Approvals | |||
|
3/6 Watts
JHM Series
xppower.com
⢠International Medical Approvals
⢠3000 VAC Reinforced Insulation
⢠60 º C Operation Without Derating
⢠2 µA Patient Leakage Current
⢠DIP 24 Package
⢠EN55011 Level A With No External Components
⢠3 Year Warranty
Specification
Input
Input Voltage Range
Input Current
Inrush Current
Input Filter
Patient Leakage
Current
Undervoltage Lockout
Input Surge
⢠12 V (10-17 VDC)
24 V (20-30 VDC)
⢠See table
⢠25 A max at 30 V
⢠Pi network
⢠2 µA max
⢠12 V model, on at <9 V, off >8.5 V
24 V model on at <18.8 V, off >16 V
⢠12 V models 25 V for 3 s
24 V model 50 V for 3 s
Output
Output Voltage
⢠See table
Output Voltage Trim ⢠±10%
Minimum Load
⢠No minimum load required
Initial Set Accuracy ⢠±1% max
Start Up Delay
⢠5 ms typical
Start Up Rise Time
⢠2 ms typical
Line Regulation
⢠±0.3% max
Load Regulation
⢠±2% max 0% to 10% load,
±1% max 10% to 100% load
Cross Regulation
⢠±4% max on dual with one output set to
50% load and the other varied from
0% to 100% load
Transient Response
⢠4% max deviation, recovery to within 1%
in <500 µs for a 50% load change at 0.25
A/µs rate
Ripple & Noise
⢠1% pk-pk max 20 MHz bandwidth
Overload Protection ⢠120% - 200%
Overvoltage Protection ⢠120% - 140%
Temperature
Coefficient
⢠±0.03/°C max
Short Circuit Protection ⢠Trip and restart (hiccup mode),
auto recovery
General
Efficiency
Isolation
Input to Output
Capacitance
Switching Frequency
Power Density
MTBF
⢠See tables
⢠5000 VAC for 10 ms (acc. to IEC60664-1),
3000 VAC reinforced isolation for 1 min.
⢠20 pF max
⢠JHM03: 180 kHz to 1.2 MHz variable
JHM06: 200 kHz to 1.6 MHz variable
⢠JHM03: 7.5 W/in3, JHM06: 17.0 W/in3
⢠>1 MHrs typical to MIL-STD-217F
at 25 °C, GB
Environmental
Operating Temperature ⢠-20 °C to +100 °C, derate from 100% load
at +60 °C to no load at 100 °C
Case Temperature
⢠+100 °C max
Storage Temperature ⢠-40 °C to +100 °C
Operating Humidity ⢠5-90%, non-condensing
Cooling
⢠Natural convection
EMC & Safety
Emissions
Immunity
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Magnetic Field
Safety Approvals
⢠EN55011 & EN55022 level A conducted &
radiated with no external components
⢠IEC60601-1-2, EN61204-3
⢠EN61000-4-2, level 2, Perf Criteria A
⢠EN61000-4-3, 10 V/m Perf Criteria A
⢠EN61000-4-4, level 3 Perf Criteria A
⢠EN61000-4-5, level 1 Perf Criteria A
⢠EN61000-4-6, 10 Vm, Perf Criteria A
⢠EN61000-4-8, 3 A/m Perf Criteria A
⢠ANSI/AMMI ES60601-1 3rd Edition
CSA-22.2 No.60601-1:2008
IEC60601-1 3rd Edition
|
▷ |