ECS65-130-SERIES Datasheet, PDF (1/5 Pages) XP Power Limited – 65/100/130 Watts | |||
|
65/100/130 Watts
ECS Series
xppower.com
⢠IT & Medical Safety Approvals
⢠65/80/100 W Convection cooled Ratings
⢠Class I & Class II Construction
⢠Industry Standard 2â x 4â Package
⢠<0.5 W No Load Input Power
⢠Low Leakage Current
⢠3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
No Load Input Power
Earth Leakage Current
Input Protection
⢠80-264 VAC (275 VAC ECS65)
(120-370 VDC), derate output power
<90 VAC, see derating curves
⢠47-400 Hz(1)
⢠65 W: 1.0/0.6 A typical at 115/230 VAC
100 W: 1.5/0.9 A typical at 115/230 VAC,
130 W: 1.9/1.1 A typical at 115/230 VAC
full load
⢠40 A max at 230 VAC, cold start at 25 ËC
⢠EN61000-3-2, class A
⢠<0.5 W
⢠65/130 W: 260 µA at 264 VAC/60 Hz max
100 W: 230 µA at 264 VAC/60 Hz max
⢠Internal T3.15/T5 A/250 V fuse in line and
neutral
Output
Output Voltage
⢠12-48 VDC (see tables)
Output Voltage Trim ⢠±10%
Initial Set Accuracy ⢠±1%
Minimum Load
⢠No minimum load required
Start Up Delay
⢠1 s typical
Start Up Rise Time
⢠50 ms
Hold Up Time
⢠16 ms min at 115 VAC
Drift
⢠±0.2% after 20 min warm up
Line Regulation
⢠±0.5% max
Load Regulation
⢠±1%
Over/Undershoot
⢠5% typical
Transient Response
⢠4% max. deviation, recovery to within
1% in 500 µs for a 50-75-50% load
change
Ripple & Noise
⢠1% pk-pk, 20 MHz bandwidth
Overvoltage Protection ⢠115-140% Vnom, recycle input to reset
Overload Protection ⢠110-160%
Short Circuit Protection ⢠Continuous trip and restart (hiccup mode)
Temperature
Coefficient
⢠0.05%/ËC
Notes
1. Safety approvals cover frequency 47-63 Hz.
General
Efficiency
Isolation
Switching Frequency
Power Density
MTBF
⢠Up to 90% model dependent
⢠4000 VAC Input to Output, 2 x MOPP,
1500 VAC Input to Ground, 1 x MOPP,
500 VDC Output to Ground, 1 x MOPP
⢠65 KHz typical
⢠65 W: 7.7 W/in3, 100 W: 10 W/in3,
130 W: 13 W/in3
⢠65 W: 850 kHrs, 100 W: 834 kHrs,
130 W: 715 kHrs to MIL-HDBK-217F
at 25 ËC, GB
Environmental
Operating Temperature ⢠-20 ËC (-40 ËC, ECS130) to +70 ËC derate
linearly from +50 ËC at 2.5%/ËC to 50%
load at +70 ËC.
Cooling
⢠Convection & fan cooled ratings (see tables)
Operating Humidity ⢠95% RH, non-condensing
Storage Temperature ⢠-40 ËC to +85 ËC
Operating Altitude
⢠3000 m
Shock
⢠30 g pk, half sine, 6 axes
Vibration
⢠2 g rms, 5 Hz to 500 Hz, 3 axes
EMC & Safety
Low Voltage PSU EMC ⢠EN61204-3, high severity level
Emissions
⢠EN55011/22 level B conducted
EN55011/22 level A radiated, level B
Suffix â-Bâ models
Harmonic Currents
⢠EN61000-3-2, class A
Voltage Flicker
⢠EN61000-3-3
Radiated Immunity
⢠EN61000-4-3, level 3 Perf Criteria A
EFT/Burst
⢠EN61000-4-4, level 3 Perf Criteria A
Surge
⢠EN61000-4-5, installation class 3
Perf Criteria A
Conducted Immunity ⢠EN61000-4-6, level 3 Perf Criteria A
Dips & Interruptions
⢠EN61000-4-11, 30% 10 ms, 60% 100 ms,
100% 5000 ms, Perf Criteria A, B, B,
EN60601-1-2, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms, Perf Criteria
A, A, A, B - 230 VAC. Consult longform
datasheet for 115 V operation.
Safety Approvals
⢠IEC60950-1 CB report, CSA 22.2
No. 60950-1, UL60950-1, TUV EN60950-1
Equipment Protection
Class
IEC60601-1 CB report, CSA 22.2
No. 60601-1, ANSI/AMMI ES60601-1,
TUV 60601-1, including risk management
⢠Class I & II (ECS65 Non â-Bâ models,
Class I only)
|
▷ |