ECP60UD01 Datasheet, PDF (1/2 Pages) XP Power Limited – 60 Watts ECP Series | |||
|
60 Watts
ECP Series
xppower.com
⢠Low Profile Design
⢠Compact Size, 4â x 2â
⢠IT & Medical Approvals
⢠Single, Dual and Triple Output
⢠<0.5 W No Load Input Power
⢠Peak Load Capability
⢠3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
No Load Input Power
Earth Leakage Current
Input Protection
⢠85-264 VAC, derate from 100% load at
90 VAC to 90% load at 85 VAC
⢠47-63 Hz
⢠0.85 A max at 115 VAC
⢠100 A max at 230 VAC, cold start at 25 °C
⢠EN61000-3-2, class A
⢠<0.5 W
⢠<250 µA at 264 VAC, 60 Hz
⢠Internal T2 A/250 V fuse in both
line and neutral
Output
Output Voltage
⢠See table
Output Voltage Trim
⢠±10% on V1 only, output 2 on multi output
models will track by the same percentage
Minimum Load
⢠10% minimum load required on all
outputs of multi output models
Start Up Delay
⢠1.3 s max
Start Up Rise Time
⢠15 ms typical
Hold Up Time
⢠16 ms min at full load at 115 VAC
Total Regulation
⢠Output 1: 2% from 0% load to 100% load
Output 2: 5% from 20% load to 100% load
Output 3: 5% from 10% load to 100% load
Remote Sense
⢠Fitted to US05 version, compensates for
0.5 V drop
Transient Response
⢠4% max deviation, recovering to less than
1% within 500 µs for 50% step load
change at 1 A/µs
Ripple & Noise
⢠1% pk-pk (2% for US05 version),
measured with 20 MHz bandwidth
Overvoltage Protection ⢠115-140% of nominal output voltage
on V1 only, recycle input to reset
Overload Protection ⢠120-160% of nominal power
Short Circuit Protection ⢠Trip and restart (hiccup mode)
Temperature
Coefficient
⢠±0.02%/°C max
General
Efficiency
Isolation
Switching Frequency
MTBF
⢠88% typical at 230 VAC and full load
⢠4000 VAC Input to Output
1500 VAC Input to Ground
500 VDC Output to Ground
⢠30-130 kHz variable
⢠>400 kHrs to MIL-HDBK-217F
at 25 °C, GB
Environmental
Operating Temperature ⢠-20 °C to +70 °C, derate from 100% load
at 50 °C to 50% load at 70 °C
Cooling
⢠Natural convection
Operating Humidity ⢠5% to 95% RH, non condensing
Operating Altitude
⢠3000 m max
Storage Temperature ⢠-40 °C to +85 °C
Shock
⢠30 g pk, half sine, 6 axes
Vibration
⢠2 g rms, 5 Hz to 500 Hz, 3 axes
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT / Burst
Surge
Conducted Immunity
Dips & Interruptions
Safety Approvals
⢠EN55011/22, level B conducted & level A
radiated
⢠EN61000-3-2, class A
⢠EN61000-3-3
⢠EN61000-4-2, ±4 kV indirect contact,
Perf Criteria A
⢠EN61000-4-3, 3 V/m, Perf Criteria A
⢠EN61000-4-4, level 3 Perf Criteria A
⢠EN61000-4-5, installation class 3,
Perf Criteria A
⢠EN61000-4-6, 3 V Perf Criteria A
⢠EN61000-4-11, 30% 10 ms, 60% 100 ms,
100% 5000 ms, Perf Criteria A, B, B
EN60601-1-2, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms,
Perf Criteria A, A, A, B
⢠EN60950-1, cUL60950-1, IEC60950-1,
EN60601-1, ANSI/AAMI ES60601-1,
CSA22.2 No.60601-1 Including Risk
Management, IEC60601-1
|
▷ |