ECP225 Datasheet, PDF (1/2 Pages) XP Power Limited – 150/225 W Convection/ Forcedcooled
225 Watts
ECP Series
xppower.com
• Low 1” Profile with 2.5” x 5” Footprint
• 150/225 W Convection/ Forced cooled
• High Efficiency up to 95%
• Medical & ITE Approvals
• Built in Fan Supply
• <0.5 W No Load Input Power
• 3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
No Load Input Power
Input Protection
• 85-264 VAC, Derate from 100% at 90 VAC
to 90% at 85 VAC
• 47-63 Hz
• 2.2 A typical at 115 VAC, 1.1 A typical at
230 VAC
• 120 A max at 230 VAC, cold start at 25 °C
• >0.95 at full load
• <230 µA at 264 VAC, 60 Hz
• <0.5 W
• Internal T3.15A/250VAC fitted in line and
neutral
Output
Output Voltage
• See tables
Output Voltage Trim • ±5%
Initial Set Accuracy • ±1% at 50 % load
Minimum Load
• No minimum load requirement
Start Up Delay
• 2 s max
Start Up Rise Time
• 55 ms typical
Hold Up Time
• 10 ms minimum at full load and 115 VAC
20 ms typical at 150 W, 13 ms typical at
225 W
Line Regulation
• ±0.5% max
Load Regulation
• ±0.5% max
Transient Response
• 4% maximum deviation, recovering to less
than 1% within 500 µs for 25% step load
Ripple & Noise
• 1% max pk-pk, 20 MHz bandwidth,
(see note 2)
Overvoltage Protection • 110% - 140% of nominal voltage on main
output. Recycle mains to reset.
Overload Protection • 110-170%
Short Circuit Protection • Trip and restart (hiccup)
Thermal Protection
• Measured internally. Auto resetting.
Temperature
Coefficient
• 0.02%/°C
Fan Supply
• 12 V at 500 mA
General
Efficiency
Isolation
Protection Level
Power Density
Switching Frequency
MTBF
• See table
• 4000 VAC Input to Output
1500 VAC Input to Ground
1500 VAC Output to Ground
• Primary to Secondary: 2 MOPP
Primary to Earth: 1 MOPP
Secondary to Earth: 1 MOPP
• 12/18 W/in3 convection/forced-cooled,
• PFC: 70-130 KHz, PWM: 50-80 KHz
• >300 kHrs to MIL-HDBK-217F at 25 °C, GB
Environmental
Operating Temperature • -20 °C to +70 °C derate from 100% load
at 50 °C to 50% load at 70 °C
Cooling
• Convection cooled: 150 W
Forced cooled: 225 W with 10 CFM
Operating Humidity • 5% to 90% RH, non condensing
Operating Altitude
• 5000 m
Storage Temperature • -40 °C to +85 °C
Shock
• IEC68-2-27, 30 g, 11 ms half sine, 3 times
in each of 6 axes
Vibration
• IEC68-2-6, 10-500 Hz, 2 g 10 mins /
sweep. 60 mins for each of 3 axes
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Dips & Interruptions
Safety Approvals
• EN55022/11, Level B conducted &
Level A radiated
• EN61000-3-2 Class A
EN61000-3-2 Class C for load >145 W
• EN61000-3-3
• EN61000-4-2, ±8 kV air, ±4 kV contact,
Perf Criteria A
• EN61000-4-3, 3 V/m, Perf Criteria A
• EN61000-4-4, level 3, Perf Criteria A
• EN61000-4-5, installation class 3,
Perf Criteria A
• EN61000-4-6, 3 V, Perf Criteria A
• EN55024, 30% 10 ms,
60%, 100 ms, 100%, 5000 ms
Perf Criteria A, A, B for high line,
A, B, B for low line at full load,
EN60601-1-2, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms,
Perf Criteria A, A, A, B for high line,
A, B, A, B for low line at full load
• UL60950-1, IEC60950-1, EN60950-1,
ANSI/AAMI ES 60601-1, IEC60601-1,
EN60601-1