CLC175_12 Datasheet, PDF (1/2 Pages) XP Power Limited – 175 Watts CLC Series | |||
|
175 Watts
CLC Series
xppower.com
⢠High Efficiency Resonant Topology
⢠Medical Safety Approvals ( M Versions)
⢠12 V Fan Output
⢠Remote Sense
⢠5V Standby Option
⢠Remote On/Off & Power Good Signal Options
⢠3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
Input Protection
⢠85-264 VAC (120-370 VDC), derate output
power 10% <90 VAC
⢠47-63 Hz
⢠3.0 A typical at 115 VAC, full load
1.5 A typical at 230 VAC, full load
⢠40 A max at 230 VAC, cold start at 25 °C
⢠EN61000-3-2, class A
⢠475 µA 264 VAC at 60 Hz, 230 µA
maximum at 264 VAC at 60 Hz (-M)
⢠Internal T5 A 250 V fuse in line
Internal T5 A 250 V fuse in line and neutral
(-M)
Output
Output Voltage
⢠12-48 V (see tables)
Output Voltage Trim ⢠±5%
Initial Set Accuracy ⢠±1% V1, ±5% V2, ±3% V3
Minimum Load
⢠0.1 A required on V1 to maintain regulation
of V2
Start Up Delay
⢠1.5 s typical, see longform datasheet for
more details
Start Up Rise Time
⢠5 ms typical
Hold Up Time
⢠16 ms min at 115 VAC
Drift
⢠±0.2% after 20 min warm up
Line Regulation
⢠±0.5% max
Load Regulation
⢠±1% V1, ±5% V2 & V3 max
Over/Undershoot
⢠5% typical
Transient Response
⢠4% max. deviation, recovery to within
1% in 500 µs for a 50-75-50% load
change
Ripple & Noise
⢠1% pk-pk V1, others 2%,
20 MHz bandwidth
Overvoltage Protection ⢠115-140% Vnom, recycle input to reset
Overload Protection ⢠110-150% V1 only
Short Circuit Protection ⢠Continuous trip and restart (hiccup mode)
Temperature
Coefficient
⢠0.05%/°C
Remote Sense
⢠Compensates for 0.5 V total voltage drop
Remote On/Off
(Inhibit/Enable)
(Option -A)
⢠Uncommitted isolated optocoupler diode,
powered diode inhibits V1 & V2
General
Efficiency
Isolation
Switching Frequency
Signals (Option -A)
MTBF
Power Density
⢠87% typical
⢠3000 VAC Input to Output
4000 VAC Input to Output (-M) 2 x MOPP
1500 VAC Input to Ground 1 x MOPP
500 VAC Output to Ground
⢠70 kHz typical
⢠Power OK - open collector,
Remote On/Off, 5 V Standby
⢠390 kHrs per MIL-HDBK-217F
at 25 °C, GB
⢠9.0 W/in3
Environmental
Operating Temperature ⢠0 °C to +70 °C derate linearly from +50 °C
at 2.5%/°C to 50% load at +70 °C
Cooling
⢠Forced-cooled >10 CFM
Operating Humidity ⢠95% RH, non-condensing
Storage Temperature ⢠-40 °C to +85 °C
Operating Altitude
⢠3000 m
Shock
⢠30 g pk, half sine, 6 axes
Vibration
⢠2 g rms, 5 Hz to 500 Hz, 3 axes
EMC & Safety
Low Voltage PSU EMC ⢠EN61204-3, high severity level
Emissions
⢠EN55011/22 level B conducted
EN55011/22 level A radiated
Harmonic Currents
⢠EN61000-3-2, class A
Voltage Flicker
⢠EN61000-3-3
Radiated Immunity
⢠EN61000-4-3, level 3 Perf Criteria A
EFT/Burst
⢠EN61000-4-4, level 3 Perf Criteria A
Surge
⢠EN61000-4-5, installation class 3
Perf Criteria A
Conducted Immunity ⢠EN61000-4-6, level 3 Perf Criteria A
Dips & Interruptions
⢠EN61000-4-11, 30% 10 ms, 60% 100 ms,
100% 5000 ms Perf Criteria A, B, B
EN60601-1, 30% 500ms, 60% 100ms,
100% 10ms, 100% 5000ms (-M)
Perf Criteria A, A (with 50% load), A, B
Safety Approvals
⢠EN60601-1, ANSI/AAMI ES60601-1,
CSA22.2 No.60601-1 per cUL, Including
Risk Management, EN60950-1,
UL60950-1
|
▷ |