AMM90PS05 Datasheet, PDF (1/2 Pages) XP Power Limited – AC-DC Power Supplies | |||
|
90 Watts
AMM90 Series
xppower.com
⢠CEC2008 & EISA2007 Compliant â¥12V
⢠Worldwide Medical Approvals
⢠Small Size/High Power Density
⢠Class I and Class II Versions
⢠Operating Temperature up to +60 °C
⢠High Efficiency
⢠Locking Output Connector â¥15 V
Specification
Input
Input Voltage
⢠90-264 VAC
Input Frequency
⢠47-63 Hz
Input Current
⢠1.2 A rms at 115 VAC,
0.6 A rms at 230 VAC
Inrush Current
⢠60 A at 115 VAC, 120 A at 230 VAC,
cold start +25 °C
Power Factor
⢠0.98 typical
Earth Leakage Current ⢠90 µA max at 115 VAC/60 Hz,
(Class I Versions)
150 µA max at 230 VAC/50 Hz
Input Protection
⢠Internal 2.5 A fuse in live & neutral lines
Output
Output Voltage
⢠5 to 48 VDC
Output Voltage Trim ⢠Not user-adjustable
Initial Set Accuracy ⢠±2 %
Minimum Load
⢠No minimum load required
Start Up Delay
⢠2 s max at 115 VAC
Start Up Rise Time
⢠<80 ms at 115 VAC
Hold Up Time
⢠15 ms minimum at full load & 115 VAC
Line Regulation
⢠0.5% maximum
Load Regulation
⢠See table
Transient Response
⢠4% max. deviation, recovery to within
1% in 500 µs for a 25% load change
Ripple & Noise
⢠2% max pk-pk (see note 1)
Overvoltage Protection ⢠112-140% Vnom, recycle input to reset
Overtemperature
Protection
⢠Unit shuts down, auto recovery
Overload Protection ⢠110-200%, auto recovery
Short Circuit Protection ⢠Trip and restart (Hiccup mode)
Temperature
Coefficient
⢠0.04%/°C
General
Efficiency
Isolation
Switching Frequency
Power Density
MTBF
⢠78% min for 5 V, 6 V & 9 V models
85% minimum for all others
⢠4000 VAC Input to Output
1500 VAC Input to Ground (Class I only)
500 VAC Output to Ground (Class I only)
⢠50-110 kHz variable
⢠3.4 W/In3
⢠150 kHrs per MIL-HDBK-217F
Environmental
Operating Temperature ⢠0 °C to +60 °C, derate from 100% power
at +40 °C to 50% power at +60 °C
Cooling
⢠Convection-cooled
Operating Humidity ⢠10-95% RH, non-condensing
Storage Temperature ⢠-20 °C to +80 °C
Operating Altitude
⢠3000 m
Shock
⢠30 g, 10 ms on 3 axes
Vibration
⢠5-100 Hz, 2.31 m/s2, 20 mins, 3 axes
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Magnetic Field
Dips & Interruptions
Safety Approvals
⢠EN55011/FCC/VCCI, Class B conducted
EN55011/FCC/VCCI, Class B radiated
⢠EN61000-3-2, Class A
⢠EN61000-3-3
⢠EN61000-4-2, level 3 Perf Criteria A
⢠EN61000-4-3, level 2 Perf Criteria A
⢠EN61000-4-4, level 3 Perf Criteria A
⢠EN61000-4-5, level 3 Perf Criteria A
⢠EN61000-4-6, 3 V Perf Criteria A
⢠EN61000-4-8, 3 A/m Perf Criteria A
⢠EN61000-4-11, 70% UT for 10 ms,
40% UT for 100 ms, <5% UT for 5000 ms
Perf Criteria A, B, B (Perf Criteria A, A, B
with 70% load)
⢠EN60601-1, UL60601-1,
CSA22.2 No. 601-1 per cUL, CE Mark
|
▷ |