Language : 

AHM Datasheet, PDF (1/6 Pages) XP Power Limited – Energy Efficiency Level V, CEC2008 AND EISA 2007
85 250 Watts
AHM Series
xppower.com
• IT & Medical Approvals
• Energy Efficiency Level V, CEC2008 & EISA 2007
• <0.5 W No Load Input Power
• Class I and Class II Models (Except AHM250)
• Very High Efficiency
• IP22 Environmental Rating
• 3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
Input Protection
No Load Input Power
• 80-264 VAC (see derating curve)
• 47-63 Hz
• 85 W: 0.8 A at 115 VAC, 0.4 A
100 W: 1.0 A at 115 VAC, 0.5 A
150 W: 1.5 A at 115 VAC, 0.75 A
180 W: 1.7 A at 115 VAC, 0.9 A
250 W: 2.3 A at 115 VAC, 0.9 A
at 230 VAC
• 70 A typical, 120 A max at 230 VAC,
cold start at 25 °C
• >0.9 at 115 VAC, 60 Hz at 250 VAC
• 85 W: 70 µA typical, 180 µA max
100 W: 70 µA typical, 180 µA max
150 W: 80 µA typical, 200 µA max
180 W: 80 µA typical, 200 µA max
250 W: 80 µA typical, 200 µA max
at 264 VAC/60 Hz
• 85 W: Internal T2.5 A/250 V
100 W: Internal T2.5 A/250 V
150 W: Internal T4.0 A/250 V
180 W: Internal T4.0 A/250 V
250 W: Internal T6.3 A/250 V
fuse in line and neutral
• <0.5 W
Output
Output Voltage
• See tables
Output Voltage Trim • Not user-adjustable
Minimum Load
• No minimum load required
Start Up Delay
• 100-300 ms typical, model dependant(1)
Hold Up Time
• 20 ms typical
Line Regulation
• ±1.5% maximum
Load Regulation
• ±3.0% maximum
Transient Response
• 5% max. deviation, recovery to within
1% in 500 µs for a 25% load change
Ripple & Noise
• Typically less than 1%, 1.5% max pk-pk(2)
Overvoltage Protection • 125% typical, Recycle input to reset
Overtemperature
Protection
• Unit shuts down, auto recovery
Overload Protection • 115-175%, auto recovery
Short Circuit Protection • Trip and restart (hiccup mode)
Temperature
Coefficient
• 0.05 %/°C
General
Efficiency
Energy Efficiency
Isolation
Switching Frequency
Power Density
MTBF
• 92% typical
• Level V
• 4000 VAC In to Out, 2 x MOPP
1500 VAC In to Earth, 1 x MOPP
500 VAC Out to Earth, 1 x MOPP
• 45-200 KHz variable
• 4.4-5.7 W/in3
• 153 kHrs typical to MIL-HDBK-217F
at 25 °C, GB
Environmental
Operating Temperature • 0 °C to +60 °C, derate linearly from 100%
power at +40 °C to 60% power at +60 °C
Cooling
• Convection-cooled
Operating Humidity • 5-95% RH, non-condensing
Storage Temperature • -40 °C to +85 °C
Operating Altitude
• 3000 m
Shock
• 3 x 30 g/11 ms shocks in both +ve & -ve
directions along the 3 orthogonal axis,
total 18 shocks.
Vibration
• Three axis 5-500 Hz at 2 g x 10 sweeps
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Magnetic Field
Dips & Interruptions
Safety Approvals
Notes
• EN55011/22, class B conducted
EN55011/22, class B radiated
• EN61000-3-2, class A
• EN61000-3-3
• EN61000-4-2, level 3, Perf Criteria A
• EN61000-4-3, level 3, Perf Criteria A
• EN61000-4-4, level 3, Perf Criteria A
• EN61000-4-5, installation class 3,
Perf Criteria A
• EN61000-4-6, level 3, Perf Criteria A
• EN61000-4-8, level 3, Perf Criteria A
• EN61000-4-11, EN60601-1-2
• EN60601-1, ANSI/AAMI ES60601-1,
CSA22.2 No. 60601-1 per cUL, Including
Risk Management, EN60950, UL60950.
Denan (PSE) certification
1. See longform datasheet for further information.
2. Measured at the end of the output cable with 10 µF electrolytic and 0.1 µF