English
Language : 

PSHBU58 Datasheet, PDF (3/4 Pages) Wall Industries,Inc. – 60 Watts AC - DC Medical Open Frame Power Supplies Single Output
Rev A
PSHBU58 SERIES
60 Watts
AC/DC Medical Open Frame Power Supplies
Single Output
NOTES
(1) Setting voltage range is a factory setting and cannot be adjusted.
(2) Output can provide up to peak load when the power supply starts up. Continuous staying in more than rated load is not allowed.
(3) In 60% rated load condition, each output is checked to be within voltage accuracy.
(4) Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
(5) Load regulation is defined by changing ±40% of measured output load from 60% rated load.
(6) Ripple & Noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47µF capacitor at rated load and
nominal line.
(7) Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated
load and nominal line.
(8) Efficiency is measured at rated load and nominal line.
DERATING CURVES
Input Voltage Derating Curve
Temperature Derating Curve
MECHANICAL DRAWINGS
PIN CHART
PIN
PSHBU58-1XX
1
RTN
2
RTN
3
OUT
4
OUT
Notes:
1.
2.
3.
Net Weight: 140g approx.
Input Connector mates with Molex housing
09-50-3031/35977-0300 and Moles
2478/35922 series crimp terminal.
Output connector mates with Molex housing
09-50-3061/35977-0600 and Molex
2478/35922 series crimp terminal.
11/10/2015
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: sales@wallindustries.com