English
Language : 

ME Datasheet, PDF (2/6 Pages) MTRONPTI – 14 pin DIP, 5.0 Volt, ECL, PECL, Clock Oscillator
Rev B
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
SPECIFICATIONS (CONTINUED)
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Nom
Max
Unit
ENVIRONMENTAL
Operating Temperature (Ambient)
Ambient
-40
+55
°C
Operating Temperature (Case)
Case
-40
+95
°C
Storage Temperature
Over Temperature Protection
Case Temperature, automatic
-40
+125
°C
107
112
117
°C
Lead Temperature
1.5mm from case for 10 seconds
260
°C
Humidity
Cooling
95
%
Free air convection
Temperature Coefficient
±0.02 ±0.05 %/°C
MTBF
Conducted EMI
MIL-HDBK-217F @ 25°C, Ground Benign
700,000 hours
EN55022 Class A
PHYSICAL
Weight
Dimensions (L x W x H)
18 grams
1.25 x 0.8 x 0.47 inches
31.8 x 20.3 x 12.0 mm
Case Material
non-conductive black plastic
Flammability
UL94V-0
MODEL SELECTION GUIDES
Model Number
ME12S5-1000W
ME12S12-500W
ME24S5-1000W
ME24S12-500W
ME48S5-1000W
ME48S12-500W
Input Voltage
12 VDC
(9 ~ 18 VDC)
24 VDC
(18 ~ 36 VDC)
48 VDC
(36 ~ 75 VDC)
Output
Voltage
5 VDC
12 VDC
5 VDC
12 VDC
5 VDC
12 VDC
SINGLE OUTPUT MODELS
Output Current
Min
Max
Input Current
Reflected
No Load Max Load Ripple Current
Output Power
Efficiency
(Typ)
Maximum
Capacitive Load
200 mA 1000 mA
30 mA
100 mA 500 mA
570 mA
641 mA
60 mA
5W
75%
1000 μF
6W
78%
470 μF
200 mA 1000 mA
20 mA
100 mA 500 mA
278 mA
313 mA
30 mA
5W
77%
1000 μF
6W
80%
470 μF
200 mA 1000 mA
10 mA
100 mA 500 mA
139 mA
156 mA
15 mA
5W
77%
1000 μF
6W
80%
470 μF
Model Number
ME12D12-250W
ME12D15-200W
ME24D12-250W
ME24D15-200W
ME48D12-250W
ME48D15-200W
DUAL OUTPUT MODELS
Input Voltage
Output
Voltage
Output Current
Min
Max
Input Current
Reflected
No Load Max Load Ripple Current
Output Power
Efficiency
(Typ)
Maximum
Capacitive Load
12 VDC ±12 VDC ±50 mA ±250 mA 30 mA
(9 ~ 18 VDC) ±15 VDC ±40 mA ±200 mA 30 mA
641 mA
641 mA
100mA
6W
78%
220 μF #
6W
78%
220 μF #
24 VDC ±12 VDC ±50 mA ±250 mA 20 mA
(18 ~ 36 VDC) ±15 VDC ±40 mA ±200 mA 20 mA
313 mA
313 mA
50mA
6W
80%
220 μF #
6W
80%
220 μF #
48 VDC ±12 VDC ±50 mA ±250 mA 10 mA
(36 ~ 75 VDC) ±15 VDC ±40 mA ±200 mA 10 mA
156 mA
156 mA
25mA
6W
80%
220 μF #
6W
80%
220 μF #
# For each output
NOTES
1. Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%.
2. The ME series requires a minimum output loading to maintain specified regulations. Operation under no-load conditions will not damage
these devices, however they may not meet all listed specifications.
3. All DC/DC converters should be externally fused at the front end for protection.
4. Other input and output voltages may be available, please contact factory.
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797