English
Language : 

PSMHBU31 Datasheet, PDF (1/2 Pages) Wall Industries,Inc. – AC/DC Medical Power Supply
Rev A
PSMHBU31 Series
Open Frame
30 Watt, Single Output
AC/DC Medical Power Supply
FEATURES
● Single Output
● RoHS Compliant
● Class II Insulation
● Internal EMI Filter
● Dimensions: 4 x 1.5 x 0.83 Inches
● Output Voltages Available from 5VDC thru 40VDC
● Wide Input Voltage Range: 90 to 264VAC, 47 to 63Hz
● Input Surge Current, Over Voltage, and Over Load Protection
DESCRIPTION
The PSMHBU31 series of AC/DC medical power supplies provides 30 Watts of continuous output power in a compact, open frame
constructed design. All models have a single output and a universal input. These power supplies meet FCC Part-18 class B and CISPR-11
EN55011 class B emission limits. They are also designed to comply with UL/cUL (UL60601-1), TUV/T-mark (EN60601-1) and new CE
requirements. This series is best suited for use in Home health care applications. All models are 100% burn in tested.
SPECIFICATIONS: PSMHBU31 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min Nom Max Unit
INPUT (Vin)
Operating Voltage Range
Input Frequency
Input Current (Low Line)
Input Current (High Line)
Inrush Current (Low Line)
Inrush Current (High Line)
Safety Ground Leakage Current
Start-Up Time
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Io = Full Load, Vin = 240VAC
Io = Full Load, Vin = 100VAC
90
264
VAC
47
63
Hz
0.8
A
0.3
0.4
A
15
18
A
25
40
A
0.1
0.3
mA
0.25
0.5
s
OUTPUT (Vo)
Output Voltage
Load Regulation
Line Regulation
Output Power
Output Current
Ripple & Noise (peak to peak)
Transient Response Time
Hold-Up Time
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
See Table
3
5
%
0.5
1
%
0
30
W
See Table
0.5
1
%
4
ms
12
18
ms
PROTECTION
Over Current Protection
110
150
%
GENERAL
Efficiency
Dielectric Withstanding Voltage
For Primary to Secondary
Io = Full Load, Vin = 230VAC
Primary to Secondary
75
85
5600
89
%
VDC
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
Temperature Coefficient
MTBF
Derate linearly from 100% Load at 40°C to 50% load at 70°C
Operating temperature at 25°C, Calculated per MIL-HDBK-217F
0
-40
5
-0.04
40
+70
+85
95
+0.04
100,000 hours
°C
°C
%
%/°C
PHYSICAL
Weight
Dimensions (L x W x H)
Approximately 7oz (200g)
4.02 x 1.50 x 0.83 inches
102.0 x 38.1 x 21.0 mm
SAFETY
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
Vin = 220VAC
Vin = 110VAC
B
Class
B
Class
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797