English
Language : 

SMFA120 Datasheet, PDF (2/2 Pages) Vitec Power – 120W Open-Frame Power Supply for Medical Equipment
Conditions
Parameter
Test Conditions
Operating Temperature
Storage Temperature
Relative Humidity
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
De-rate linearly from 100% load at 50°C to 50% load at 70°C
Min.
Typ.
Max.
Unit
0
50
70
°C
-40
85
°C
5
95
%
0.1M
Hrs
Approvals and Compliances
Parameter
Dielectric Withstanding Voltage for Primary to secondary
Dielectric Withstanding Voltage for Primary to Ground
Isolation Resistance
EMI requirements for CISPR-11
EMI requirements for FCC PART-18
Test Conditions
Primary to secondary
Primary to ground
Test Voltage = 500VDC
Vin=220VAC
Vin=110VAC
Mechanical and PIN out
Min.
5656
2828
50
B
B
Unit
VDC
VDC
MΩ
CLASS
CLASS
PIN CHART
MODEL
PIN 1 2 3 4 5 6 7 8 9 10 11 12
SMFA120-SXX-12pin OUT OUT OUT OUT OUT OUT RTN RTN RTN RTN RTN RTN
MODEL
PIN 1 2 3 4 5 6
SMFA120-SXX-6pin OUT OUT OUT RTN RTN RTN
Note:
1.Dimensions are shown in mm.
2.Weight: 420g approx.
3.Input connector mates with Molex housing
09-50-3051 and Molex 2478 series crimp terminal.
4.Output Connector mates with screw terminal (Terminal
Block)(16-22AWG)or Molex housing 09-50-3121 and Molex
2478 series crimp terminal