English
Language : 

SMDA63 Datasheet, PDF (2/2 Pages) Vitec Power – 63W Desk Top Power Supply for Medical Equipment
Conditions
Parameter
Test Conditions
Operating Temperature
Storage Temperature
Relative Humidity
Operation temperature at 25℃, calculated per MIL-HDBK-217F
Derate linearly from 100% load at 50°C to 50% load at 70°C
Min.
0
-40
5
0.1M
Approvals and Compliances
Parameter
Test Conditions
Dielectric Withstanding Voltage for Primary to secondary
Primary to secondary
Dielectric Withstanding Voltage for Primary to Ground
Primary to ground
Isolation Resistance
Test Voltage = 500VDC
EMI requirements for CISPR-11
Vin=220VAC
EMI requirements for FCC PART-18
Vin=110VAC
Mechanical and PIN out
Typ.
Max.
Unit
50
70
°C
85
°C
95
%
Hrs
Min.
5656
2828
50
B
B
Unit
VDC
VDC
MΩ
CLASS
CLASS
Note:
1. Dimensions are shown in mm &
inch
2. Weight: approx. 510-560g
(Exclude the input cord)
3. Optional output connector.